51
Ventilation continues.
The oxygen inlet pressure is more
than the specified level.
High O2 Pressure
If the Air source is available,
ventilation continues with Air only.
The safety valve opens if the pressure
of both Air and Oxygen sources drops
or disconnects.
The oxygen inlet pressure is lower
than the specified level.
This alarm will not be enabled if the
oxygen concentration is set to 21%.
Low O2 Pressure
Ventilation continues.
The air inlet pressure is more than
the specified level.
High Air Pressure
If the Oxygen source is available,
ventilation continues with Oxygen
only.
The safety valve opens if the pressure
of both Air and Oxygen sources drops
or disconnects.
The air inlet pressure is lower than
the specified level.
This alarm will not be enabled if the
oxygen concentration is set to 100%.
Low Air Pressure
Ventilation continues.
If the measured gas temperature
parameter (gas temp) exceeds the
allowed limit (45 ° C).
High Gas
Temperature
Ventilation continues.
--
Error code 17
Ventilation continues.
--
Error code 2
Ventilation continues with the
following settings: Mode: PCV, Rate:
15, Ti: 1, PEEP: 2, Pcontrol: 20,
Trigger: Pressure, Ptrig: -3, O2: 100
--
Fatal Error 5
Ventilation continues with the
following setting: Mode: PCV, Rate:
15, Ti: 1, PEEP: 2, Pcontrol: 15,
Trigger: Pressure, Ptrig: -3, O2: 100
--
Fatal Error 6
Ventilator reaction to alarm
Alarm Definition
Medium Priority
Alarm
Ventilation continues.
Respiratory rate (including
spontaneous and mandatory breaths)
has exceeded high alarm setting in
three respiratory cycles.
High Rate
Ventilation continues.
Respiratory rate (including
spontaneous and mandatory breaths)
has violated low alarm setting in
three respiratory cycles.
Low Rate
The inhalation phase will switch to
the exhalation phase (whether alarm
is activated or not) once Vti has
exceeded alarm limit.
Inspired tidal volume has exceeded
high alarm setting in three
consecutive respiratory cycles.
Vti Limit Reached
Ventilation continues.
Exhaled tidal volume has exceeded
high Vt
e
alarm setting in three
consecutive respiratory cycles.
High Vte