11
When the device is turned on, audible and visual alarms will be enabled for a few
seconds to check the alarm system. Do not use the device if integrity of the alarm
system is in doubt and contact the Customer Service department.
Use only sterilized tracheal tube.
Prolonged use of tracheal tube may harm to patient.
Due to leaks around the mask, the patient expiratory volume may be different from
the measured expiratory volume in NIV mode.
Before using the device, prepare necessary equipment to monitor expiratory CO2 in
NIV mode.
There could be hazard of electrical shock by opening the ventilator casing. All
servicing and future upgrading to this equipment must be carried out by personnel
trained and authorized by manufacturer.
Only medically pure oxygen which is oil-free should be used for ventilation. Do not
use anesthetics and explosive gases.
To prevent fire hazard, keep all ignition sources away from the system.
Ensure that an alternate source of ventilation is always available.
In case of uncertainty of device performance, replace the respina ventilator with
another ventilator or ventilation device.
If the ventilator is damaged and ventilation fails, stop using the ventilator because it
may lead to the patient death.
To avoid risk of electric shock, the device must be connected to a mains supply with
proper grounding.
When the ventilator is connected to an external device, the power cable should be
used to ensure proper grounding.
To prevent EMC effect on the ventilator, the device should not be used adjacent to or
in conjunction with other equipment and if adjacent use is necessary, the device
should be observed to verify normal operation in the configuration in which it will be
used.
Do not use cellular phone in the vicinity of this equipment. High level of
electromagnetic radiation emitted from such devices may result in strong interference
with the ventilator performance.
Do not replace any accessories or other parts of the device during the patient
ventilation.
Adding attachments to breathing system may decrease pressure and affect the
ventilator function.
Before using the ventilator ensure
that a battery with minimum charge is installed.
If you’re going to store the device for an extended time, charge the battery and
remove it from the device.
Do not use the device in case of the battery failure.
If the oxygen sensor is faulty or uncalibrated, measurement values will not be
reliable.