Technical Specifications
Technical Specifications
Ventilation
Modes
Assisted Control Mandatory Ventilation
A/C
Synchronized Intermittent Mandatory Ventilation
SIMV
Spontaneous Ventilation
SPONT
Breath Types
Volume–controlled breaths
VCV, V–SIMV (With Dual control
in the breath)
Pressure–controlled breaths
PCV, P–SIMV, PSV
Volume Targeted Pressure–controlled breaths
(Pressure Regulated Volume Control) & (Volume
Support)
PRVC–CMV, PRVC–SIMV, VSV
Dual level PEEP breaths
APRV (With Auto Tlow)
Patient Type
Patient Height
42 – 250 cm
IBW (Ideal Body Weight )
5 – 200 Kg
Gender
Male, Female
VBS (Ventilator
Breathing
System)
Compliance
0.5 – 5 ml/cmH2O
Inspiratory Resistance
0 – 2.5 cmH2O.s/l
Expiratory Resistance
0 – 2.5 cmH2O.s/l
Apnea Backup
Settings
On (Rate, Pcontrol, Ti); OFF
Breath triggering
Pressure triggering
(-2) – (-20) cmH2O
Flow triggering
2 – 20 l/min
Additional
Setting
Respiratory Rate
Accuracy: ±1 b/min or 4% wig
2– 150 b/min
Respiratory Rate apnea
Accuracy:±1 b/min or 4% wig
5 – 150 b/min
Tidal Volume
Accuracy:
For 50–2000 ml: ± (4 ml + 15% Actual reading)
For 20–49 ml: ± (7 ml + 15% Actual reading)
20 – 2600 ml
(Compliance & BTPS compensated)
PEEP / CPAP
Accuracy: ±12% or ±(4% Actual reading)
wig
0 – 50 cmH2O
Pcontrol
Accuracy: ±12% or ±(4% Actual reading)
wig
2 – 80 cmH2O
Psupport
Accuracy: ±12% or ±(4% Actual reading)
wig
0 – 80 cmH2O
Peak Flow
Accuracy: ±12% or ±(3 l/min +4% Actual reading)
wig
0.2 – 250 l/min
Flow Pattern
Square, Decel.
NIV
Non- Invasive Ventilation
On or Off;
Available in P-SIMV, PSV and apnea
modes.
Leak Comp
Automatic Leak Compensation
On or Off
I–Time (Ti)
Accuracy: ± 0.02 sec
0.1 – 10.0 sec