The power supply alarms include:
Ventilator reaction to alarm
Alarm Definition
High Priority
Alarm
Ventilation continues.
Internal battery voltage is lower than
20%.
Battery Low
If the device is connected to the mains
power, ventilation continues.
The internal battery is not available.
No Battery
If the device is connected to the mains
power, ventilation continues.
The battery failure.
Battery Deffect
Ventilator reaction to alarm
Alarm Definition
Medium Priority
Alarm
Ventilation continues.
Internal battery voltage is lower than
20- 40%.
Battery Low
Ventilation continues by using
internal battery.
The device is unplugged.
AC Unplug
Definition
Technical messages
Inspiratory Hold maneuver has been enabled.
Insp.Hold
Expiratory Hold maneuver has been enabled.
Exp.Hold
Manual Ispiratory maneuver has been enabled.
Manual Insp.
100% O2 is delivered to the patient.
O2 100%
Nebulizer has been enabled.
Nebulizer
The suction maneuver has been enabled.
Suction
Apnea backup has been disabled in PSV, VSV and APRV modes.
Apnea off
Warning
Announcing Oxygen related alarms have a delay about 40 seconds.
The disconnection alarm may not be activated if the endotracheal tube is
disconnected from the patient.