93
Pause (insp plateau)
Accuracy: ±0.02 sec
0 – 75% of Inspiratory time
Rise time
Fast, Med, Slow
Exhalation sensitivity
Accuracy: ±12% or ±(3 l/min +4% Actual reading)
wig
5 – 85 % of peak flow
Oxygen (FiO2)
Accuracy (Delivery): ± (3 %) full scale
21 – 100 %
Phigh
Accuracy:±12% or ±(4% Actual reading)
wig
5 – 50 cmH2O
Plow
Accuracy: ±12% or ±(4% Actual reading)
wig
0 – 47 cmH2O
Thigh
Accuracy: ±0.02 sec
1.0 –29.8 sec
Tlow
Accuracy: ±0.02 sec
0.2 –22.0 sec
Auto Tlow
Accuracy: ±12% or ±(3 l/min +4% Actual reading)
wig
5 – 85 % of peak flow
Nebulizer
Accuracy: ±10 s
On or Off;
User selected duration (1– 480
minutes)
Humidification Type
None, HME, Humidifier
Compliance Compensation
On or Off
Graph Setting
Waveforms: flow, pressure, volume
Loops: P/V, F/V
Trend Data
all parameters, 72 hours
Trend resolution
6 seconds
Displayed Measurements
6 parameters
Audio/ Settings
Audio Level
Monitored /
Displayed Patient
Values
Pressure Values
Ppeak (peak pressure during a breath)
Accuracy: ±12% or ±(4% Actual reading)
wig
0 – 120 cmH2O
PEEP (pressure at end exhalation)
Accuracy: ±12% or ±(4% Actual reading)
wig
0 – 120 cmH2O
Pmean (averaged mean pressure)
Accuracy: ±12% or ±(4% Actual reading)
wig
0 – 120 cmH2O
P
endInsp
Accuracy: ±12% or ±(4% Actual reading)
wig
0 – 120 cmH2O
Volume / Flow Values
Vte (expiratory tidal volume)
Accuracy:
For 50- 2600 ml: ±(4 ml + 15% Actual reading)
For 20- 49 ml: ±(7 ml + 15% Actual reading)
0 – 3000 ml
Vti
Accuracy:
For 50- 2600 ml: ±(4 ml + 15% Actual reading)
For 20- 49 ml: ±(7 ml + 15% Actual
0 – 3000 ml