34
Figure 5-9 Trend
In each curve, the vertical axis represents the numerical value of the parameter and the
horizontal axis of the time. it is possible to change time frame by clicking on the written time
in the middle of horizontal axis area. These time frames are as follows 60 min, 6 h, 12 h, 24
h, 36 h and 72 h. It should be noted that the changing of the time frame applied on a curve is
also applied simultaneously on two other curves.
it is also possible to change the scale of
parameter by clicking on the vertical axis.
The parameters that could be trended are listed in below table:
Mechanics
Advanced
Basic/ Fundamental
R insp
Pplateau
Vti
R exp
PEEP
Vte
C stat
Auto PEEP
MV
C dyn
Ti
MV Spont
WOBimp
Te
Rate
RSBI
Ti/T tot
Rate Spont
RCe
Leak
P peak
Gas Temp
2
O
PendInsp
P mean
PFe
5-7
Alarms
Selecting Alarms will load Alarms Menu in which you can set alarm limits.
There is also an
Event log option on this page, which is described following. Set alarm limits in two ways:
1.
Click on alarm limits (high and low) and set your desired value.
2.
Select AutoSet key to set alarm limits of all parameters except Ppeak, Leak, Apnea
time, Vt lim automatically based on setting values which are listed below: