Sample tube types
Most commonly used brands/types.
Closed tubes with concentric cap only.
Barcode reader:
Barcode reader type
CCD.
Reading capabilities
Most common barcode labels
Code39, ITF, Industrial 2 or 5, CodaBar,EAN/UPC, CODE128
Environment:
Sound level
Less than 65 dBA
Environment temperature
18 - 28 °C
Relative humidity
10-90%
Total power specifications
Mains voltage
115/230VAC, 50/60 Hz
Fuse (20 x 5 mm)
Per ESR Unit
Slow blow 230V 2.5 Amp / 115V 5 Amp
Transport Unit
Slow blow 230V 1.6 Amp / 115V 3.15 Amp
XN1
XN2
XN3
Power consumption
Standby
90 VA
2x90VA
3x90VA
Maximum
650 VA
2x1000 VA
3x1500 VA
Heat output
Standby
86 Watt
2x86 Watt
3x86 Watt
Operation
460 Watt
2x460 Watt
3x460 Watt
Dimensions
XN1 FRL
XN2 FRL
XN3 FRL
Width
1400 mm
2500 mm
3600 mm
Height
1530 mm
1530 mm
1530 mm
Depth
1150 mm
1150 mm
1150 mm
Total Weight (empty)
238 kg
519 kg
740 kg