E9: Air flow failure!
ESR analyser unit was not able to
get a stable reading during the
vacuum test before aspiration the
sample.
•
Check for leakage on the
pipette or fill nozzle.
E10: Sample probe was jammed.
Check both needles before
sampling!
Sample probe was probably
jammed when going down and
exceeded the maximum current
level.
Sample probe went back to its
home position after the error.
•
Check if outer needle is
clogged up with rubber.
•
Sample probe maybe bend.
E11: Sample probe not on position
(going down)! Timeout error!
Sample probe did not go down
within a certain time limit.
•
Sample probe motor is faulty.
•
Sample probe motor driver on
needle board is faulty.
•
Sample probe is blocked.
E12: Dilution error: wrong or no
diluent flow. Check the diluter!
Diluter malfunction
•
Check diluent flow sensor
•
Check tubes diluter system
E13: Fillnozzle not in home
position!
Fill nozzle did not reach the Home
position within a certain time limit.
•
Fill nozzle motor is faulty.
•
Fill nozzle motor driver on
nozzle board is faulty.
•
Fill nozzle is blocked.
E14: Outer needle motor position
error! (home)
Outer needle did not reach the
home (top) sensor within a certain
time limit.
•
Check home (top) sensor.
•
Outer needle motor is faulty.
•
Outer needle motor driver on
needle board is faulty.
•
Outer needle is blocked.
E18: Carousel position error!
Check Rinse position.
Value of potentiometer does not
match the value stored in memory
of the current rinse position.
•
Check if the rinse position is
right.
•
Set correct rinse position and
perform a "Learn carousel
position".
•
Check mechanical connection
potentiometer.
E19: Drive motor timeout!
Drive motor did not move or motor
is blocked
•
Check the home sensor
•
Motor is faulty
•
Motor driver on drive board is
faulty
E22: Waste bottle full!
Empty waste bottle and clear error. •
Check level sensor.
E23: "Fill " sensor out of range.
Check/clean this sensor !
The Fill sensor has reached a
critical level.
Continuing could result in filling
errors.
•
Check and/or clean the Fill
sensor. If cleaning fails, replace
sensor