Wash each pipette:
When the pipette belt turns one position, the pipette at the rinse position will be rinsed and dried,
regardless if it was filled or not.
▪
Wash all pipettes:
Before executing this function, check carefully if there are samples in the pipette belt that need to be
measured.
Any remaining samples will be washed away and will
NOT
be measured!
All pipettes on the pipette belt are washed and dried once.
A warning is shown on the display: <Pipette data will be lost!>.
▪
Wash only sample pipettes:
Before executing this function, check carefully if there are samples in the pipette belt that need to be
measured. Any remaining samples will be washed away and will NOT be measured!
All pipettes which are currently holding samples are washed and dried once.
A warning is shown on the display: <Pipette data will be lost!>.
▪
Fill & Clean:
This button starts the Fill & Clean procedure. During prolonged use of the instrument, proteins are
building up in the Westergren pipettes which need to be removed using a strong cleaning agent. This
function fills all pipettes with a cleaning agent and removes the cleaning agent after a specified time.
▪
End-of-day wash:
All pipettes will be washed once and needle, fill-nozzle and rinse-nozzle (wash station) are primed.
5.5.1.1 - Fill and Clean screen
Fill & Clean:
Automatic fill and clean function, each individual pipette on pipette belt will be filled with cleaning solution.
During prolonged use of the instrument, proteins are building up in the Westergren pipettes which need to
be removed using a strong cleaning agent.
This cycle takes about 90 minutes.
The Fill & Clean function is part of the monthly maintenance procedure.
A warning is shown on the display: <Pipette data will be lost!>.
By toggling the switch ON the Fill and clean adapter is used.
By toggling the switch OFF the Fill and clean without adapter is used.
See chapter Maintenance
for more information.
Starrsed program
MRN-164_2-EN Version 2.02 Interrliner