C. If applicable replace the bacterial HEPA filter (For Maintenance Level 4: Exchange filter QWLV040002)
On waste bottle (If used):
Exchange bacterial filter
QWLV040001
on the waste bottle assembly.
Exchange Rinse and Saline tube assembly
New rinse pump tube assembly
ESRI090902
.
New saline pump tube assembly
ESRI090903
New tube replacement:
1. Open left cover.
2. Pull pump tube slightly downwards and at the same time towards the front of the unit to release the tube out
of the pump plate holder.
3. Remove the old tube from the peristaltic pump rotor.
4. Disconnect the tubing at both ends of the tube connectors.
5. Connect new tubing to both ends of the connectors.
6. Place one end of the tube in the pump plate holder.
7. Pull the new tube over the peristaltic pump rotor.
8. Pull pump tube slightly downwards and at the same time towards the back of the Interrliner.
Work instruction Interrliner
MRN-164_2-EN Version 2.02 Interrliner
115