Result is out of range, the applicable values for the
acceptable range depend on the user setting. E116 is
shown in the status line of the Sample screen and the QC
icon is blinking on the Sample screen.
ESR Result is given.
▪
Try new QC sample tube (normal samples will be
finished)
▪
Check acceptable range in QC settings. If results are
continuously out of range but the statistics show
identical/stable results, it should be considered to
expand the acceptable assay range with QC Settings
▪
If this error persists check/clean instrument
E117: Uncorrected QC result is out of acceptable range, but corrected result is within range!
ESR Result is given.
Temperature correction not activated.
▪
Consider QC Sample as correct. The mean value is
assayed with temperature correction
▪
Check temperature correction setting.
E118: Uncorrected QC result is within acceptable range, but corrected result is out of range!
ESR Result is given.
Temperature correction not activated.
▪
Consider QC Sample as not correct
▪
Try new QC sample tube (normal samples will be
finished)
▪
Check acceptable range in QC settings
▪
If this error persists check/clean instrument
▪
Check temperature correction setting.
QC result with ESR error
no ESR Result is given
▪
Check general ESR data, see
▪
Check sample tube volume
▪
Try new QC sample tube
QC result with ESR warning
ESR Result is given
▪
Check general ESR data, general
▪
Check limit settings
Note: QC Error messages are only shown and stored in QC results and not send to LIMS.
QC result is given with the same general errors and warnings as a normal patient ESR-result
Screen messages
QC icon is blinking at Sample screen
The last QC sample was not within acceptable range or
has no result
▪
Press on QC icon
▪ Press "Accept" to continue sampling without
performing a new QC, continuing could produce
incorrect results.
▪ Press "Cancel" to return. Try new QC sample tube
(normal samples will be finished)
Troubleshooting
MRN-164_2-EN Version 2.02 Interrliner
81