7.1.2 - Report 30 Minute mode
+ REPORT E
(Not to scale)
- StaRRsed--
Date 25/04/2019
Time:
15:28
1
2
3
4
5
6
7
8
9
10
915001
42
84
75
CLEAR
17
30
23
EDTA
7.2 - ESR Error
Error messages can be found on the printout in column 10.
If errors are found during the measurement, the ESR analyser unit will give an audible alarm.
The Error message is displayed on the main screen.
7.2.1 - ESR Error and Warning code messages
ESR "ERROR" and "WARNING" code messages. This code appears in the "sample data record" at column 10.
Error: No ESR result is given
Warning: Result should be reviewed before release.
The following codes are defined:
0
No errors
1
No cells/plasma found
Error
No contents could be detected in the pipette.
2
ESR Probably > 140 mm
Error
Extremely high ESR value.
3
Too many borders found
Error
More than three borders found, possibly air
bubbles. See Section Troubleshooting
.
4
Column height <nnn>
Warning
Column height must be between 180 and
210mm. <nnn> = the actual column height.
5
Measure error
Warning
The down count is not equal to the up count
from the measure head.
6
Bubbles on top
Warning
Air bubbles on top of the ESR. See Section
7
Limit error
Error
One of the following limits are out of the setting
range:
•
ESR Time
•
Column height
•
Dilution
•
Bubbles on top
•
Hazy aspect
•
Temperature
Reporting
64
MRN-164_2-EN Version 2.02 Interrliner