E181: Barcode rotator up / down
motor timeout!
The Barcode vertical motor didn’t
reach its up or down position
within a certain time limit.
▪
Check up and down sensors.
▪
Faulty Barcode vertical motor.
▪
Barcode rotator unit
mechanically blocked.
▪
Broken wire to barcode
vertical motor.
▪
Faulty Barcode vertical motor
driver on Indexer board.
E182:
E183: Indexer motor timeout!
The indexer didn’t reach its
position within a certain time limit.
▪
Fuse was triggered.
▪
Faulty indexer motor.
▪
Indexer mechanically
blocked.
▪
Broken wire(s) to indexer
E184: Indexer error: barcode
rotator not in up position!
The indexer could not move to a
new position, because the barcode
rotator was not in its up position.
▪
Check Barcode rotator up /
down motor.
▪
Check sensors of Barcode
rotator up / down unit.
E185: Indexer error: elevator
platform not in back position!
The indexer could not move to a
new position, because the elevator
platform was not in its back
position.
▪
Check platform slider motor.
▪
Check home sensor of
platform unit.
E193: Elevator motor timeout!
The elevator didn’t reach its
position within a certain time limit.
▪
Fuse was triggered.
▪
Faulty lift motor.
▪
Elevator mechanically
blocked.
▪
Broken wire(s) to lift motor.
▪
Faulty lift motor driver on
Elevator board.
E194: Tube unit front / back motor
timeout!
The platform slider didn’t reach the
front or back position within a
certain time limit.
▪
Check platform home and
front sensors.
▪
Faulty platform slider motor.
▪
Platform unit mechanically
blocked.
▪
Broken wire to platform slider
motor.
▪
Faulty platform slider motor
driver on Elevator board.
Appendix for Interrliner
MRN-164_2-EN Version 2.02 Interrliner
101