Fill and clean
Starrsed Cleaning Agent preparation Interrliner ESR analyser unit: Fill
and clean:
This cycle takes about 90 minutes.
1. Fill the clean adapter EHST110907 with hot de-ionized water. (+/-
150 ml, 80 °C)
2. Add 15 ml Starrsed Cleaning Agent (QRR 010905) to the hot
water in the adapter.
3. Place the cap on the adapter and mix well.
4. Put the adapter with cleaning solution on the lower tube holder.
5. Select
Maintenance
tab,
Prime/Clean
, button
Fill and clean
.
Start Fill and clean procedure:
1. Select button OK.
2. The needle goes down and the process is started.
3. When all the pipettes are filled, the needle goes back to the home position.
4. After the Fill & clean process is finished, a notification is given to stop the procedure.
5. Remove the adapter from the needle assembly and press "Stop".
Work instruction Interrliner
116
MRN-164_2-EN Version 2.02 Interrliner