E195: Sample unit in / out motor

timeout!

The Sample arm unit in / out didn’t

reach the in (= needle) or out (=

rack) position within a certain time

limit.

Check sample arm out and

home sensors.

Faulty rotation motor.

Sample arm unit mechanically

blocked.

Broken wire to rotation motor.

Faulty rotation motor driver

on Elevator board.

E196: Mixer motor timeout!

The Mixer didn’t reach the up or

down position within a certain time

limit.

Check mixer sensors.

Faulty mixer motor.

Mixer unit mechanically

blocked.

Broken wire to mixer motor.

Faulty mixer motor driver on

Elevator board.

E197: Sample unit in / out error:

elevator platform not in back

position!

The Sample arm unit could not

move to a new position, because

the elevator platform was not in its

back position.

Check platform slider motor.

Check back sensor of the

platform slider unit.

E198: Sample unit in / out error:

elevator not in up position!

The Sample arm unit could not

move to a new position, because

the elevator was not in its up

position.

Check lift motor (see also

rack transport error 63).

E199: Tube unit front / back error:

sample arm not in out position!

The platform slider unit could not

move to a new position, because

the sample arm was not in its out

position (= rack position).

Check rotation motor.

Check out sensor of Sample

arm.

E200: Sample unit in / out and

Tube unit front / back not on

position during initialisation of the

elevator!

The initialisation routine of the

elevator was cancelled, because

the Sample arm unit was not in its

out position (= rack position) and

the Platform slider unit was not in

its front position. The positions of

both units are unknown!

Check rotation motor.

Check out sensors of sample

arm unit.

Check platform slider motor.

Check front sensor of

platform slider unit.

E201: Outer needle not in up

position during initialisation of the

elevator!

The initialisation routine of the

elevator was cancelled, because

the outer needle (sample tube

holder) was not in its up position.

Check outer needle up sensor.

Check outer needle motor.

E210: Bad communication

between ESR-unit and Input-pool!

Bad communication between ESR

analyser unit and Input-pool, or

the communication between the

ESR analyser unit and the Input-

pool was interrupted.

Switching off the Interrliner

could cause this error. This is

normal.

Check RS232 cable between

Elevator board and Interrliner

Transport board.

Appendix for Interrliner

102

MRN-164_2-EN Version 2.02 Interrliner

Содержание EHST109621

Страница 1: ...Instructions for Use Interrliner MRN 164_2 EN Version 2 02...

Страница 2: ......

Страница 3: ...ciple 13 2 Instrument description 14 2 1 PC Operation and User Interface 15 2 2 Technical specifications 16 2 3 Used reagents 19 2 3 1 Starrsed Rinse solution 19 2 3 2 Starrsed Saline 19 2 3 3 Starrse...

Страница 4: ...ple history QC 41 5 3 5 6 Linked QC IDs 41 5 3 5 7 QC Result analysis 42 5 3 6 Set start end date 43 5 3 7 History sample analyse option 43 5 3 8 History analyse results 45 5 4 Reagents screen 46 5 4...

Страница 5: ...2 Turn off sequence 61 7 Reporting 62 7 1 Result Printout 62 7 1 1 Report 60 Minute mode 63 7 1 2 Report 30 Minute mode 64 7 2 ESR Error 64 7 2 1 ESR Error and Warning code messages 64 7 3 Limit erro...

Страница 6: ...Fill time out error 75 11 8 Hazy reports 76 11 9 Leaking pipettes 77 11 10 Liquid level sensor not sensing 77 11 11 Air bubbles 77 11 11 1 Foam in column 78 11 11 2 Pipette looks like zebra crossing 7...

Страница 7: ...acement 88 13 3 4 Fill and clean procedure 88 13 4 Level 3 maintenance 90 13 5 Check or replace sample probe or outer needle 91 13 6 Reagents installation 92 13 7 Cleaning Touch screen display 93 14 A...

Страница 8: ...al MRN 164_2 is applicable for the following instruments EHST109621 Interrliner XN1 FRL EHST109622 Interrliner XN2 FRL EHST109623 Interrliner XN3 FRL with Software version V6 xx Application 8 MRN 164_...

Страница 9: ...andatory H Schavemaker 2 01 jan 2020 Update and publication H Schavemaker 2 00 June 2017 Instrument description Safety Program Maintenance Technical specifications additions Explanation of symbols Int...

Страница 10: ...e tubes reduces the possibility of contamination for the user and environment Standard glass Westergren pipettes are used in which the measurement can be corrected to a constant temperature 18 C Celsi...

Страница 11: ...Service manual MRN 125 1 4 Explanation of symbols The following symbols can be used in this manual and for the instrument Warning sign to prevent personal injury due to biohazard Warning sign to preve...

Страница 12: ...vailable in several languages 2 User Manual UM If available Intended for the lab supervisor Contains information from the IFU and additional information concerning settings service higher maintenance...

Страница 13: ...of a sample 1 9 Dilution principle The principle of adding Diluent to a flow of whole blood is unique The Interrliner has the capability of monitoring the air displacement during the aspiration cycle...

Страница 14: ...acing successively each sample tube on the ESR analyser unit for aspiration Conveyor belt running from the Start pool Front Return line for transport of racks back to Sysmex HST XN after sampling Prot...

Страница 15: ...m The user interface is intuitive and can be activated via the keyboard or the optional touch screen All the data from each sample including the raw measuring data and a pictorial representation of th...

Страница 16: ...ical Laboratory World 1994 Allowed blood specimen types For EDTA mode Whole blood with 1 EDTA anticoagulant Automatic dilution 4 vols blood 1 vol Starrsed Diluent Reported result mm after 1 hour Throu...

Страница 17: ...0 60 Hz Fuse 20 x 5 mm Per ESR Unit Slow blow 230V 2 5 Amp 115V 5 Amp Transport Unit Slow blow 230V 1 6 Amp 115V 3 15 Amp XN1 XN2 XN3 Power consumption Standby 90 VA 2x90VA 3x90VA Maximum 650 VA 2x100...

Страница 18: ...22 kg double 26 triple LIS connectivity Communication Ethernet and serial bi directional according various protocols Data storage Storage medium Hard disk on external PC Storage capacity indication a...

Страница 19: ...blood samples in Starrsed ESR analyzers Approximately 0 5 ml Diluent is used for each sample Approximately 2 5 ml is used for one Prime cycle The diluent container is a 5 liter container QRR 010931 2...

Страница 20: ...n the cabinet beneath the Start Pool Attention The sample robot unit is powered by the ESR analyzers power supply Switching the transport units OFF will not switch OFF the sample robot A central power...

Страница 21: ...ment must be made inoperative and be secured against any unintended operation The matter should then be referred to qualified technicians Safety protection is likely to be impaired if for example the...

Страница 22: ...cover is closed and locked To get access to the tube handling mechanism e g for maintenance or other necessary operator intervention Switch the Sample mode OFF by pressing the sample button The sampl...

Страница 23: ...creen The Reagents screen is used to check the reagent levels and log reagent replacement To activate priming sequences and cleaning operations the screen Maintenance is used The Settings and Service...

Страница 24: ...www rrmechatronics com Login part The following program description is valid for software from version 6 12 1 and up and up 5 2 Sample screen The main menu is displayed during operation To access oth...

Страница 25: ...DE the instrument uses the standard saved settings For instance the ESR time is set back to 60 minutes or 30 minutes according the used method The User Manual button is at the bottom of the menu Click...

Страница 26: ...r this sample ESR is required and waiting to be done Sample was aspirated successfully Sample was aspirated but with fill errors After processing the rack the information of the rack is transferred to...

Страница 27: ...us Information about the current status of the instrument is shown here such as the selected mode EDTA or Citrate selected method 60 or 30 minute and symbols that draw attention to certain maintenance...

Страница 28: ...5 2 3 Pipette information The following information is shown Starrsed program 28 MRN 164_2 EN Version 2 02 Interrliner...

Страница 29: ...te is marked to be washed when it reaches the rinse station This indicator is also on when a sample could not be aspirated properly Pipette disabled In case of a failing missing or broken pipette indi...

Страница 30: ...lected result to the printer Print result header Only if option Settings General settings Print after measurement is switched ON it is possible to print a result header Send all to HOST Send all resul...

Страница 31: ...process time After measuring the pipette the pipette data is transferred to the sample history files In the header of the table the names of the columns are shown Double click the header of any column...

Страница 32: ...a selected time frame of the day Show results of a number of past days Default value is set for 7 days Show results of a specific day Show results of the range between the first selected date to the...

Страница 33: ...ys Show results of a specific day Show results of the range between the first selected date to the next selected date From the selected Sample ID detailed information is shown on this screen Sample ID...

Страница 34: ...r Pipette in which the sample was measured Error code Shows any ESR error or warning code e g Too many borders found A complete overview of measurement data is shown with activating the option Display...

Страница 35: ...ervice General Barcode the rack number is shown under date time The Clear rack history button will clear the contents of the rack history file and restart to build up a new rack history file 5 3 3 1 D...

Страница 36: ...Bubbles on top Increasing samples with bubbles indicate the need for maintenance of the aspiration system see Foam in column ESR errors Increasing ESR errors may indicate the need for maintenance see...

Страница 37: ...sted in Patient results Note This part of the software can only be used in combination with Starrsed Control as quality control material 5 3 5 1 QC Normal results table The results from Starrsed Contr...

Страница 38: ...sed Control expiry date is exceeded it is not possible to continue with this QC sample The sample is not measured but the failed attempt is logged in the table Expected ESR This is the temperature cor...

Страница 39: ...MS is activated at QC Settings Related patient results This screen is simular to the Display sample history screen The background colour of the patient history table is switched to light yellow to dis...

Страница 40: ...viation the average deviation of all QC results compared with the expected ESR Coefficient of variation ratio of the standard deviation to the expected ESR expressed in a percentage Number of QC resul...

Страница 41: ...g QC result The results are presented in the layout of the Display sample history screen Depending on the frequency of QC samples related patient results may span over multiple days and are listed per...

Страница 42: ...Settings the links can be removed automatically when a usable ESR result has been reported for this particular Lab ID 5 3 5 7 QC Result analysis Authorized staff should identify and differentiate acc...

Страница 43: ...5 3 6 Set start end date Enter the Start End date and time 5 3 7 History sample analyse option Starrsed program MRN 164_2 EN Version 2 02 Interrliner 43...

Страница 44: ...today s results between start time and end time or make a selection for 1 A specific number of past days 2 A specific date 3 A range of days from start date to end date Starrsed program 44 MRN 164_2...

Страница 45: ...all the samples with a dilution rate 110 are displayed in the table By selecting Dilution 90 all the samples with a dilution rate 90 are displayed in the table Aspect By selecting one of the three Haz...

Страница 46: ...o fill in all reagent information If not filled in correctly the instrument will give an alarm and does not aspirate samples The article number always has to be added this to prevent replacing a reage...

Страница 47: ...ory of the used reagents Select the reagent type on the right side For external use of the information all the logged reagent data can by exported to EXCEL CSV format by clicking Export to Excel CSV S...

Страница 48: ...me for all reagents Input screen for new reagents Make a selection to add new default setting or delete the current information and continue with Next 5 4 2 1 New reagent input cont Data can be entere...

Страница 49: ...he date when the reagent was placed 5 Check if the preview box shows the correct information then press OK 5 5 Maintenance screen When there is a sensor alarm an alarm indicator is shown in the tab Th...

Страница 50: ...outer needle sample probe and fill nozzle are washed Prime Diluter The Diluter prime cycle is 5 strokes of the syringe Prime Nozzle with water After each aspiration the fill nozzle is flushed with de...

Страница 51: ...onged use of the instrument proteins are building up in the Westergren pipettes which need to be removed using a strong cleaning agent This function fills all pipettes with a cleaning agent and remove...

Страница 52: ...er does not start during the aspiration the status of this sensor must be checked Click the Check button the OK sign is shown if the sensor value is in range Check Diluent flow sensor This sensor is o...

Страница 53: ...position correct Measure head start position wrong When a test pipette is installed at the measuring position the result of the test pipette is displayed in the field ESR mm When the sensor is out of...

Страница 54: ...of the instrument and locate possible problems This log can be saved e g to a memory stick by clicking button Save As Note since software version 6 12 5 and up is a sample counter included in every e...

Страница 55: ...an alarm indicator is shown in the tab Maintenance All performed maintenance functions are logged automatically This log can be saved e g to a memory stick by clicking button Save As 5 5 5 Maintenance...

Страница 56: ...art to wash all pipettes needle fill nozzle and rinse nozzle wash station The function can be set up for automatic execution in the following screen Close program will only close down the program 5 5...

Страница 57: ...restart sample mode when finished if needed 5 5 8 End of day wash schedule settings Select the time of the day in hours and minutes for automatic start of this function 5 6 Settings screen The Settin...

Страница 58: ...1 2 Start pool Ensure that racks are correctly placed in the Start Pool the groove on the side of the rack must catch the strip at the right side of the storage area Note Do not push racks manually a...

Страница 59: ...es of the Diluter 4 Prime Nozzle with water activates the fill nozzle water valve De ionized water must flow through the tube connected to the fill nozzle cap 5 Prime Waste system activates the disinf...

Страница 60: ...he Interrliner may remain switched ON continuously However the customer should consider environmental issues such as energy consumption when the instrument is not to be used for some time It is also r...

Страница 61: ...equence Close the Interrliner software Switch the Interrliner OFF Switch optional printer OFF Switch the ESR analyser unit OFF Switch the Interrliner rack system OFF Operation MRN 164_2 EN Version 2 0...

Страница 62: ...U 9 Compact bidirectional 10 Compact unidirectional String format for StaRRsed 11 StaRRsed III V14 12 Vesmatic 13 Sedimatic 15 14 Sedimatic 100 15 Opus bidirectional 16 Advia 120 bidirectional 17 Advi...

Страница 63: ...21 60 23 EDTA 905006 5 5 CLEAR 22 60 23 EDTA 079 905007 24 60 23 Too many borders found 905008 25 60 23 L_err 84 75 200 EDTA 905002 905003 905004 Sample results with hazy aspect 905005 Sample result...

Страница 64: ...e following codes are defined 0 No errors 1 No cells plasma found Error No contents could be detected in the pipette 2 ESR Probably 140 mm Error Extremely high ESR value 3 Too many borders found Error...

Страница 65: ...nute method and temperature correction 75 with a correct column height L_err 84 75 200 means 84 mm in the 60 minute method and temperature correction 75 with a correct column height Note If limit erro...

Страница 66: ...see page 131 reports are shown when the change from the hazy level to the cell plasma separation level occurs outside the given distance The following code messages are reported printed in column 5 Th...

Страница 67: ...as a standard Starrsed Control should not be used past the expiry date Mechatronics as supplier of the Starrsed Control shall not be liable for any claimed damages arising from other than intended us...

Страница 68: ...value and range and the expiry date The history of QC results is maintained internally Error messages are generated when the QC results are outside the acceptable range QC samples can be optionally re...

Страница 69: ...es Do not mix residual material with material from other tubes Do not re use empty tubes The software interface is described in the chapter History screen Starrsed Control should be disposed of as med...

Страница 70: ...6 2 QC Result analysis Authorized staff should identify and differentiate acceptable unacceptable random errors and trends and or shifts in systematic errors from the statistical data Depending on the...

Страница 71: ...must be met 1 Waste tube must not exceed 5 meters or 18 feet in length 2 Drain height must not be higher than the original waste container inside the instrument Disclaimer Check the specifications of...

Страница 72: ...entation time 60 or 30 min may be exceeded However the start time is saved and therefore the actual sedimentation time can be checked Important system settings are kept in an internal Flash Eprom insi...

Страница 73: ...his section can usually not be solved by the operator Refer to the Service manual for more information available only in English The error numbers are displayed in the PC software 11 1 Errors 183 and...

Страница 74: ...bes between the containers and pumps valves Unscrew the cap from the bottle or cubitainer Check the pick up tubes in the bottle or cubitainer Check the tubes for blockages or kinks 11 5 Reagents alarm...

Страница 75: ...ontainer blocked Replace the tube Waste pump failure Exchange the waste pump cassette If the error returns call for service Electrical bridge between the waste level electrodes Clean waste separator s...

Страница 76: ...growth of micro organisms in the diluter system It is extremely important that the system is kept sterile Erroneous haziness looks more like haemolytic plasma than normal hazy plasma If there is an a...

Страница 77: ...ps of Saline to the De ionized water to increase the conductivity 11 11 Air bubbles After a normal aspiration the Westergren pipette must be free of air bubbles In the following examples different pat...

Страница 78: ...e fill nozzle washer or O ring 3 Check for air in diluter system 4 Check that the sample probe O ring is not leaking 5 Check transparent mixing piece block for cracks 11 11 2 Pipette looks like zebra...

Страница 79: ...op sensor by more than 10mm One air bubble can result in ESR Error 3 11 11 4 One air bubble rising in pipette 1 Usually this is caused by a wet or dirty fill nozzle The blood column should not reach r...

Страница 80: ...le tube Small air bubbles result in ESR Error 3 11 11 6 Random air bubbles in pipette 1 Check Diluent flow by priming the diluter system 2 Insufficient sample volume Random air bubbles result in ESR E...

Страница 81: ...t of range ESR Result is given Temperature correction not activated Consider QC Sample as not correct Try new QC sample tube normal samples will be finished Check acceptable range in QC settings If th...

Страница 82: ...cceptable range The last QC result should be evaluated by authorized staff to decide whether the Interrliner may run patient samples depending on the the nature of errors Press Yes to continue samplin...

Страница 83: ...Compare only results from one batch If Lab ID is used check the linked Starrsed Control ID It is possible that a new batch is in use without changing to the new assayed mean value Systematic QC error...

Страница 84: ...echanical obstructions and remove these 2 Clear the error by pressing F3 on the keyboard of the Input pool When the error occurs again switch OFF all units and call for service 12 2 Errors 33 and 37 T...

Страница 85: ...nstruction Daily WI 257 Daily Weekly WI 258 Weekly maintenance Level 4 maintenance WI 1215 Level 4 maintenance Every 7500 samples Level 3 maintenance WI 224 Level 3 maintenance Every 23000 samples Lev...

Страница 86: ...Fill stop sensor box Fill stop sensor FS 90 140 165 Diluter Start sensor check Go to tab Maintenance Check sensor Select Diluter start sensor box Diluter start sensor 400 550 700 Measure sensor check...

Страница 87: ...acement New rinse pump tube assembly ESRI090902 New saline pump tube assembly ESRI090903 New tube replacement 1 Open left cover 2 Pull pump tube slightly downwards and at the same time towards the fro...

Страница 88: ...ivided in the column Vacuum stabilisation errors may occur 13 3 4 Fill and clean procedure Note Each pipette on the pipette belt will be filled with Starrsed Cleaning Agent after one hour the first pi...

Страница 89: ...ed the needle goes back to the home position 4 After the Fill clean process is finished a notification is given to stop the procedure 5 Remove the adapter from the needle assembly and press Stop The i...

Страница 90: ...over the sample probe 7 Press Continue 8 The fill and clean process is started 13 4 Level 3 maintenance Level 3 maintenance is level 4 maintenance adding the following extra s 1 Replace the pinch val...

Страница 91: ...the sample probe or outer needle Lift the robotic protection hood Needle exchange 1 Unscrew sample probe manually 2 Mark each tube for easier reconnecting to the correct nipple 3 Disconnect the tubes...

Страница 92: ...ing door and place the containers on the turntable 2 Remove the cubitainer screw caps and pull the necks of the bottle packs out of the cardboard box 3 Install the level sensors and spacers 4 Add new...

Страница 93: ...ter After each reagents change the fluid system must be primed 1 Select Maintenance Prime Clean 2 Perform the applicable prime step to fill the relevant tubes with reagent and remove air 13 7 Cleaning...

Страница 94: ...14 Appendix for Interrliner Appendix section Appendix for Interrliner 94 MRN 164_2 EN Version 2 02 Interrliner...

Страница 95: ...Sample probe motor is faulty Sample probe motor driver on needle board is faulty Sample probe is blocked E5 Duplicated ID Sample rejected Sample already in carousel Wait until sample is measured Check...

Страница 96: ...nozzle motor is faulty Fill nozzle motor driver on nozzle board is faulty Fill nozzle is blocked E14 Outer needle motor position error home Outer needle did not reach the home top sensor within a cer...

Страница 97: ...setting Selected result path is not valid Software is using the default setting Check result path setting Check if network or USB devices are used E30 No ACK NACK received from host after sending inqu...

Страница 98: ...osition Could not start the position motor because the outer needle or sample probe is not its home position top Check outer needle home sensor Check sample probe home sensor Faulty Outer needle motor...

Страница 99: ...sum error Indexer settings After reading the settings from EEPROM the calculated checksum did not match with the checksum stored in EEPROM Default Indexer settings are loaded Faulty EEPROM 24C01 on El...

Страница 100: ...ted Bad I2C cable s Note the boards in the ESR analyser unit are connected in series E169 I2C communication error between PC and Drive board There was a communication error between the Driver board an...

Страница 101: ...n motor Check sensors of Barcode rotator up down unit E185 Indexer error elevator platform not in back position The indexer could not move to a new position because the elevator platform was not in it...

Страница 102: ...tform slider unit could not move to a new position because the sample arm was not in its out position rack position Check rotation motor Check out sensor of Sample arm E200 Sample unit in out and Tube...

Страница 103: ...k the InteRRliner I2C cable s E213 Bad communication between ESR unit and Input pool The ESR analyser unit did not receive a good answer second part after sending an inquiry to the Input pool Check RS...

Страница 104: ...End of day wash X X X X X X Clean outside aspiration needle X X X X X X Check tubing diluent syringe X X X X X X Clean outside instrument X X X X X X Clean Fill nozzle X X X X X Clean waste separator...

Страница 105: ...Replace Sample probe assembly X ESRI050909 1 Replace Pipette valves bodies 84 pieces X QTST040001 84 Replace Outer needle assembly X ESRI050950 1 Replace all tubing with Tubing set X EHST079200 1 Com...

Страница 106: ...SRI070924 Fill tube assembly to regulator ESRI070924 Fill tube assembly from regulator QWLV030021 Diluter Syringe QWLV080002 Quad Ring Appendix for Interrliner 106 MRN 164_2 EN Version 2 02 Interrline...

Страница 107: ...15 Work instruction Interrliner Work instruction section Work instruction Interrliner MRN 164_2 EN Version 2 02 Interrliner 107...

Страница 108: ...bes and connections for leaks Check that liquid does not run back after the pumps have stopped 3 Clean the outer needle with disinfectant 4 Check tubing from the syringe for trapped air bubbles 5 Chec...

Страница 109: ...liner Revision 001 2019 Clean Fill nozzle 1 2 3 4 Clean fill nozzle The use of a toothbrush and detergent is recommended Carefully scrub the fill nozzle inner part Use a tissue to dry the fill nozzle...

Страница 110: ...read of the glass jar the top rim of the glass jar the O ring for the HEPA filter and the black PVC pipe on the back Replacing C If applicable replace the bacterial HEPA filter For Maintenance Level 4...

Страница 111: ...nance Check sensor Select Check Diluent flow sensor box Press test When test is finished signal Down and signal Up must be green Separator check Go to tab Maintenance Check sensor Select Check Separat...

Страница 112: ...Prime Clean and perform the Prime Diluent Diluter function 7 Wipe outer surface and stainless steel plate below the pipettes with disinfectant 8 Check or replace if needed two drip trays ESRI010318 W...

Страница 113: ...ner Revision 003 2019 Clean Fill nozzle and exchange O ring Fill Nozzle 1 2 3 4 5 6 Clean fill nozzle The use of a toothbrush and detergent is recommended Carefully scrub the fill nozzle inner part Us...

Страница 114: ...all parts with hot water and a brush B Use some acid free vaseline on the screw thread of the glass jar the top rim of the glass jar the O ring for the HEPA filter and the black PVC pipe on the back...

Страница 115: ...p tube slightly downwards and at the same time towards the front of the unit to release the tube out of the pump plate holder 3 Remove the old tube from the peristaltic pump rotor 4 Disconnect the tub...

Страница 116: ...l 4 Put the adapter with cleaning solution on the lower tube holder 5 Select Maintenance tab Prime Clean button Fill and clean Start Fill and clean procedure 1 Select button OK 2 The needle goes down...

Страница 117: ...sor check Go to tab Maintenance Check sensor Select Check Diluent flow sensor box Press test When test is finished signal Down and signal Up must be green Separator check Go to tab Maintenance Check s...

Страница 118: ...Prime Clean and perform the Prime Diluent Diluter function 7 Wipe outer surface and stainless steel plate below the pipettes with disinfectant 8 Check or replace if needed two drip trays ESRI010318 W...

Страница 119: ...ner Revision 004 2019 Clean Fill nozzle and exchange O ring Fill Nozzle 1 2 3 4 5 6 Clean fill nozzle The use of a toothbrush and detergent is recommended Carefully scrub the fill nozzle inner part Us...

Страница 120: ...ssembly ESRI090903 New tube replacement 1 Open left cover 2 Pull pump tube slightly downwards and at the same time towards the front of the unit to release the tube out of the pump plate holder 3 Remo...

Страница 121: ...it Fill and clean This cycle takes about 90 minutes 1 Fill the clean adapter EHST110907 with hot de ionized water 150 ml 80 C 2 Add 15 ml Starrsed Cleaning Agent QRR 010905 to the hot water in the ada...

Страница 122: ...Check sensor Select Check Fill stop sensor box Fill stop sensor FS 90 140 165 Diluter Start sensor check Go to tab Maintenance Check sensor Select Diluter start sensor box Diluter start sensor 400 550...

Страница 123: ...ube with use of the lever 3 Pull the waste separator towards the front of the ESR analyser unit Note The separator has two sensor connectors at the rear 4 Remove bacterial HEPA filter 5 Fill waste sep...

Страница 124: ...place 7 Remove the protection caps on from the tubes 8 Connect the two tubes from the waste pump cassette on the luers at the steel plate under the waste pump Exchange Pinch valve tube Replace the pin...

Страница 125: ...trapped air bubbles 6 If trapped air bubbles are found go to tab Maintenance click button Prime Clean and perform the Prime Diluent Diluter function 7 Wipe outer surface and stainless steel plate bel...

Страница 126: ...disinfectant 6 Leave the disinfectant in the system for 15 minutes Step 2 1 Take the diluent suction tube out of the disinfectant 2 Wipe the tube clean and dry with a tissue 3 Empty the bottle and ref...

Страница 127: ...l parts with hot water and a brush B Use some acid free vaseline on the screw thread of the glass jar the top rim of the glass jar the O ring for the HEPA filter and the black PVC pipe on the back Rep...

Страница 128: ...s cycle takes about 90 minutes 1 Fill the clean adapter EHST110907 with hot de ionized water 150 ml 80 C 2 Add 15 ml Starrsed Cleaning Agent QRR 010905 to the hot water in the adapter 3 Place the cap...

Страница 129: ...e needle goes back to the home position 4 After the Fill clean process is finished a notification is given to stop the procedure 5 Remove the adapter from the needle assembly and press Stop The instru...

Страница 130: ...edle 4 Pull the sample probe complete with outer needle towards the front edge of the ESR analyser unit The outer needle must be supported to protect it from falling 5 Slide the new sample probe into...

Страница 131: ...of sedimentation setting of erythrocytes red blood cells in a blood column during a specified time Hazy A sedimentation is reported to be hazy when the boundary between blood plasma and erythrocytes...

Страница 132: ...f day wash procedure 60 52 End of day wash schedule settings 57 EQAS 12 Errors 165 171 and 210 214 73 Errors 181 185 and 193 201 74 Errors 183 and 193 73 Errors 24 26 38 42 43 62 67 70 75 80 84 Errors...

Страница 133: ...rror messages 69 QC Normal results table 37 QC Result analysis 70 42 QC Results 69 QC Results screens 36 Quality control 67 Quality control procedure 68 Quality control troubleshooting 80 Quick start...

Страница 134: ...U Usage options 68 Used reagents 19 W Waste disposal 71 Weekly 86 Work instruction Interrliner 107 134 MRN 164_2 EN Version 2 02 Interrliner...

Отзывы: