E195: Sample unit in / out motor
timeout!
The Sample arm unit in / out didn’t
reach the in (= needle) or out (=
rack) position within a certain time
limit.
▪
Check sample arm out and
home sensors.
▪
Faulty rotation motor.
▪
Sample arm unit mechanically
blocked.
▪
Broken wire to rotation motor.
▪
Faulty rotation motor driver
on Elevator board.
E196: Mixer motor timeout!
The Mixer didn’t reach the up or
down position within a certain time
limit.
▪
Check mixer sensors.
▪
Faulty mixer motor.
▪
Mixer unit mechanically
blocked.
▪
Broken wire to mixer motor.
▪
Faulty mixer motor driver on
Elevator board.
E197: Sample unit in / out error:
elevator platform not in back
position!
The Sample arm unit could not
move to a new position, because
the elevator platform was not in its
back position.
▪
Check platform slider motor.
▪
Check back sensor of the
platform slider unit.
E198: Sample unit in / out error:
elevator not in up position!
The Sample arm unit could not
move to a new position, because
the elevator was not in its up
position.
•
Check lift motor (see also
rack transport error 63).
E199: Tube unit front / back error:
sample arm not in out position!
The platform slider unit could not
move to a new position, because
the sample arm was not in its out
position (= rack position).
▪
Check rotation motor.
▪
Check out sensor of Sample
arm.
E200: Sample unit in / out and
Tube unit front / back not on
position during initialisation of the
elevator!
The initialisation routine of the
elevator was cancelled, because
the Sample arm unit was not in its
out position (= rack position) and
the Platform slider unit was not in
its front position. The positions of
both units are unknown!
▪
Check rotation motor.
▪
Check out sensors of sample
arm unit.
▪
Check platform slider motor.
▪
Check front sensor of
platform slider unit.
E201: Outer needle not in up
position during initialisation of the
elevator!
The initialisation routine of the
elevator was cancelled, because
the outer needle (sample tube
holder) was not in its up position.
▪
Check outer needle up sensor.
▪
Check outer needle motor.
E210: Bad communication
between ESR-unit and Input-pool!
Bad communication between ESR
analyser unit and Input-pool, or
the communication between the
ESR analyser unit and the Input-
pool was interrupted.
▪
Switching off the Interrliner
could cause this error. This is
normal.
▪
Check RS232 cable between
Elevator board and Interrliner
Transport board.