Clean waste separator and exchange Waste cassette assembly
Be careful, as there may be blood in the cassette. First, make up some disinfectant and put this in the
waste separator. Press
Prime Disinfectant
to pump disinfectant through the pump cassette.
Clean Separator
The waste system must be cleaned before replacing the waste
pump cassette.
1. Open the left cover and remove the waste container. The
waste separator is now visible.
2. Lift the stainless steel vacuum tube with use of the lever.
3. Pull the waste separator towards the front of the ESR
analyser unit. (Note: The separator has two sensor connectors
at the rear)
4. Remove bacterial HEPA filter.
5. Fill waste separator with 100ml disinfectant or 100 ml water
with 2% bleach.
6. Replace bacterial HEPA filter.
7. Lift left cover.
8. Lift stainless steel vacuum tube up.
9. Insert the waste separator sliding it over the support shelf.
10. Push the waste separator towards the rear, with the sensor
connectors in the holes.
11. Release the stainless steel vacuum tube.
12. Replace the waste container.
13. Close left cover.
Work instruction Interrliner
MRN-164_2-EN Version 2.02 Interrliner
123