6. Place one end of the tube in the pump plate holder.
7. Pull the new tube over the peristaltic pump rotor.
8. Pull pump tube slightly downwards and at the same time towards the back of the Interrliner.
Fill and clean
Starrsed Cleaning Agent preparation Interrliner ESR analyser unit: Fill
and clean:
This cycle takes about 90 minutes.
1. Fill the clean adapter EHST110907 with hot de-ionized water. (+/-
150 ml, 80 °C)
2. Add 15 ml Starrsed Cleaning Agent (QRR 010905) to the hot
water in the adapter.
3. Place the cap on the adapter and mix well.
4. Put the adapter with cleaning solution on the lower tube holder.
5. Select
Maintenance
tab,
Prime/Clean
, button
Fill and clean
.
Start Fill and clean procedure:
Work instruction Interrliner
MRN-164_2-EN Version 2.02 Interrliner