Reference Manual
00809-0100-4738, Rev DA
May 2006
Rosemount 3095FB
5-5
Disassembly Procedures
Read the following information carefully before disassembling a transmitter.
General information concerning the process sensor body and electrical
housing is included in the following sections.
20409 15
A
Sensor module is NOT
updating
The sensor module has undergone a component or software failure.
Replace the sensor module as described on page 5-7. Contact your
Field Service Center.
20409
14
A
Sensor board eeprom not
initialized, default values used
The sensor electronics has not initialized properly. Replace the sensor
module as described on page 5-7. Contact your Field Service Center.
20409
13
A
Sensor microprocessor does
not respond
The sensor module has undergone a component or software failure.
Replace the sensor module as described on page 5-7. Contact your
Field Service Center.
20409
12
A
Sensor board eeprom burn
failure
The sensor module has undergone a component or software failure.
Replace the sensor module as described on page 5-7. Contact your
Field Service Center.
20409
11
A
Sensor hardware
incompatible with software
The sensor hardware is incompatible with the software. Replace the
sensor module as described on page 5-7. Contact your Field Service
Center.
20409
10
A
Sensor CRC error (static
region)
Sensor CRC checksum failed indicating corrupted sensor module
memory. Replace the sensor module as described on page 5-7.
20409
9
A
Sensor CRC error (dynamic
region)
Sensor CRC checksum failed indicating corrupted sensor module
memory. Replace the sensor module as described on page 5-7.
20409
8
Reserved
NA.
20409
7
Reserved
NA.
20409
6
Flash output board eeprom
soft
(recoverable) error
If this is a common error, replace the output electronics board as
described on page 5-7.
20409 5
A
Flash output board eeprom
hard
(non-recoverable) error
The transmitter electronics has undergone a component or software
failure. Replace the output electronics board as described on
page 5-7.
20409
4
Flash output board eeprom
time out
If this is a common error, replace the output electronics board as
described on page 5-7.
20409
3
Reserved
NA.
20409
2
A
Non-volatile database CRC
error
Non-volatile database checksum failed. Attempt to restore non-volatile
database to factory defaults (see page 41 in Modbus Master
Document). If unable to restore these factory defaults, replace the
output electronics board as described on page 5-7.
20409
1
Write protect status flag
Reflects write protect jumper position. No action required.
20409
0
Bit not used
Bit not used
TABLE 5-2. Rosemount 3095FB Alarm and Event Summary
Floating
Point Reg./
Bit Posit.
Sets
Alarm/
Warning
Flag?
Alarm
Description
Action
Explosions can result in death or serious injury. Do not remove the instrument cover
in explosive environments.
Содержание 3095FB
Страница 2: ......
Страница 4: ......
Страница 8: ...Reference Manual 00809 0100 4738 Rev DA May 2006 Rosemount 3095FB 1 2 ...
Страница 144: ...Reference Manual 00809 0100 4738 Rev DA May 2006 Rosemount 3095FB 5 12 ...
Страница 168: ...Reference Manual 00809 0100 4738 Rev DA May 2006 Rosemount 3095FB C 4 ...