Reference Manual
00809-0100-4738, Rev DA
May 2006
Rosemount 3095FB
5-4
20407
0
A
SP sensor shorted
The sensor module has undergone a component or software failure.
Replace the sensor module as described on page 5-7. Contact your
Field Service Center.
20408
15
A
SP signal is unreasonable -
open bridge
This display means that the transmitter absolute (or gage) pressure
reading exceeds its sensor limits. There are two possible causes.
Either the transmitter is overpressured, or it has a sensor malfunction.
Check the pressure input to the transmitter. If an overpressure
condition exists, correct it. If not, replace the sensor module as
described on page 5-7.
20408
14
A
PT signal exceeded Upper
Range Limit + 10%
Check the transmitter RTD connector and RTD screw terminals to
ensure the RTD cable is properly connected. Verify that the process
temperature is between -40F and 400F. If process temperature
exceeds these limits, correct the temperature. If process temperature
is within these limits, replace the sensor module as described on
page 5-7.
20408
13
W
PT signal exceeded Upper
Range Limit
No action required.
20408
12
W
PT signal exceeded Upper
Operating Limit
No action required.
20408 11
PT signal is less than Lower
Operating Limit
No action required.
20408
10
W
PT signal is less than Lower
Range Limit
No action required.
20408
9
A
PT signal is less than Lower
Range Limit - 10%
Check the transmitter RTD connector and RTD screw terminals to
ensure the RTD cable is properly connected. Verify that the process
temperature is between -40 °F and 400 °F. If transmitter temperature
exceeds these limits, correct the temperature. If transmitter
temperature is within these limits, replace the sensor module as
described on page 5-7.
20408
8
A
RTD is disconnected
Check the transmitter RTD connector and RTD screw terminals to
ensure the RTD cable is properly connected.
20408
7
A
ST signal is greater than high
limit
This message indicates that the ambient temperature limit of the
transmitter is being exceeded. Verify that the transmitter ambient
temperature is between -40 °F and185 °F. If transmitter temperature
exceeds these limits, correct the temperature. If transmitter
temperature is within these limits, replace the sensor module as
described on page 5-7.
20408
6
A
ST signal is less than lower
limit
This message indicates that the ambient temperature limit of the
transmitter is being exceeded. Verify that the transmitter ambient
temperature is between -40F and 185F. If transmitter temperature
exceeds these limits, correct the temperature. If transmitter
temperature is within these limits, replace the sensor module as
described on page 5-7.
20408
5
Reserved
NA.
20408
4
Reserved
NA.
20408
3
Reserved
NA.
20408
2
Reserved
NA.
20408
1
Reserved
NA.
20408
0
Reserved
NA.
TABLE 5-2. Rosemount 3095FB Alarm and Event Summary
Floating
Point Reg./
Bit Posit.
Sets
Alarm/
Warning
Flag?
Alarm
Description
Action
Содержание 3095FB
Страница 2: ......
Страница 4: ......
Страница 8: ...Reference Manual 00809 0100 4738 Rev DA May 2006 Rosemount 3095FB 1 2 ...
Страница 144: ...Reference Manual 00809 0100 4738 Rev DA May 2006 Rosemount 3095FB 5 12 ...
Страница 168: ...Reference Manual 00809 0100 4738 Rev DA May 2006 Rosemount 3095FB C 4 ...