![Rosemount 3095FB Скачать руководство пользователя страница 12](http://html.mh-extra.com/html/rosemount/3095fb/3095fb_reference-manual_1487388012.webp)
Reference Manual
00809-0100-4738, Rev DA
May 2006
Rosemount 3095FB
2-4
Review Installation
Considerations
When choosing an installation location and position, take into account the
need for access to the transmitter. For dimensional drawing information see
page A-6.
Process Flange Orientation
Mount the process flanges with sufficient clearance for process connections.
For safety reasons, place the drain/vent valves so the process fluid is directed
away from you when the vents are used. In addition, consider the possible
need for a testing or calibration input.
Housing Rotation
See “Consider Housing Rotation” on page 2-11.
Terminal Side of Electronics Housing
Mount the transmitter so that the terminal side is accessible. A 0.75-inch (19
mm) clearance is required for cover removal. Use a conduit plug on the
unused side of the conduit opening.
Circuit Side of Electronics Housing
Provide 0.75 inches (19 mm) clearance if possible for cover removal. Three
inches of clearance is required for cover removal if a display is installed.
Cover Installation
Always install the electronics housing covers metal-to-metal to ensure a
proper seal.