2.
Loosen the housing rotation set screw with a hex wrench, and back
off one full turn (see Figure 5-3).
Figure 5-4. Loosening the
Housing Rotation Screws
3.
Unscrew the housing from the module, making sure the shroud and
sensor cable do not catch on the housing. Damage can occur to the
cable if the internal shroud and sensor cable rotate with the housing.
Carefully pull the shroud and sensor ribbon cable assembly through
the housing opening.
REASSEMBLY
PROCEDURE
Follow these procedures carefully to ensure proper reassembly:
Before removing the sensor module from the electrical housing, disconnect the
electronics board power cable from the sensor module. This will prevent damage to
the sensor module ribbon cable.