Model 3095FB Multivariable Transmitter with Modbus Protocol
Modbus Protocol Guide
Page 76 of 79
Appendix C: Modbus Mapping of 32-bit Floating Point Registers
32-bit Floating Point Registers
Modbus Access
DESCRIPTION
Address
Address
Type
Attributes
Functional
Area
Data / Control
7401
FP register
RO
PVs
DP
7402
FP register
RO
SP
7403
FP register
RO
PT
7404
FP register
RO
Diagnos-
tics
Status Bytes
7405
FP register
RO
Status Bytes
7406
FP register
RO
Status Bytes
7407
FP register
RO
Sensor lim-
its
DP upper range limit
7408
FP register
RO
DP lower range limit
7409
FP register
R/W, WP
DP upper operating limit
7410
FP register
R/W, WP
DP lower operating limit
7411
FP register
RO
SP upper range limit
7412
FP register
RO
SP lower range limit
7413
FP register
R/W, WP
SP upper operating limit
7414
FP register
R/W, WP
SP lower operating limit
7415
FP register
RO
PT upper range limit
7416
FP register
RO
PT lower range limit
7417
FP register
R/W, WP
PT upper operating limit
7418
FP register
R/W, WP
PT lower operating limit
7419
FP register
R/W, WP
DP calibra-
tion
DP offset
7420
FP register
R/W, WP
DP slope
7421
FP register
R/W, WP
DP damping