Getting Started
R&S
®
Spectrum Rider FPH
67
User Manual 1321.1011.02 ─ 11
1. In the "Instrument Setup" dialog box, select frequency clock or reference level
"Align" item to perform alignment.
2. The R&S Spectrum Rider displays the "Perform Alignment" dialog.
3. Select "Yes" to overwrite existing default alignment data and proceed to perform
alignment. The R&S Spectrum Rider displays the "Alignment" dialog.
Depending on the selected alignment item, different input signal is required. See
.
Figure 3-7: Input signal for frequency clock alignment
Figure 3-8: Input signal for reference level alignment
4. Connect an input signal of 1.1 GHz, -10dBm at the "RF IN" port for frequency align-
ment or 100 MHz, -20 dBm at the "RF IN" port for reference level alignment.
5. Select "Continue" to start the internal alignment on the instrument.
The R&S Spectrum Rider overwrites existing adjustment data and performs the
internal alignment.
If wrong input signal is detected, the R&S Spectrum Rider displays a message of a
failed alignment dialog.