Receiver Mode
R&S
®
Spectrum Rider FPH
284
User Manual 1321.1011.02 ─ 11
1. Press [AMPT] key.
2. Select "Unit" softkey.
The R&S Spectrum Rider opens a submenu that contains the possible units for
selection.
3. Select the unit.
Defining the bargraph scale
The scale of the bargraph is defined by the reference level and level range.
The reference level is the maximum power level that the bargraph displays. You should
set the reference level in a way that the signal level does not exceed the reference
level and high enough so that the signal does not disappear in the inherent noise.
1. Press [AMPT] key.
2. Select "Ref Level" softkey.
The R&S Spectrum Rider opens an input field to define the reference level.
3. Define the reference level that best fits the measurement.
In the default state, the reference level corresponds to the level at the right of the
bargraph.
The reference level position is shown by a triangle in the bargraph scale label.
Bargraph icon
Descriptions
Blue bar
Current power level.
Maximum hold power level.
Red bar
The red bar shows the limit line level.
If the limit line is violated, the color changes to a
brighter red indicator.
Triangle
Reference power level.
You can move the reference position to another position on the scale.
1. Press [AMPT] key.
2. Select "Range / RefPos" softkey.
The R&S Spectrum Rider opens a submenu that contains the range and reference
position for selection.
3. Select the "Ref Position" menu item.
The R&S Spectrum Rider opens an input field to define the reference position.
4. Enter the number of the grid line you want the reference level to be at.