Spectrum Analyzer Mode
R&S
®
Spectrum Rider FPH
126
User Manual 1321.1011.02 ─ 11
Figure 6-1: Screen layout for channel power measurements
1 = Measurement mode
2 = Channel power numerical
3 = Measurement standard
4 = Channel bandwidth
5 = Channel bandwidth graphical (blue lines)
6 =
7 =
8 = Channel power measurement softkey menu
6.1.2.1
Selecting the Standard
If you need to perform measurements that are conformed to a telecommunications
standard, you can activate one of the predefined standards that are already stored in
the R&S Spectrum Rider memory. However, you can create new configurations to per-
form measurements on other standards as well.
1. Press [MEAS] key.
2. Select "Standard" softkey.
The R&S Spectrum Rider opens a file manager dialog box to select the standard.
3. Select one of the available standards.
The R&S Spectrum Rider loads the configuration of the selected standard. It auto-
matically sets the optimal span, resolution bandwidth, video bandwidth, sweep time
and detector for the standard.
You can create and edit standards with the R&S InstrumentView software and transfer
them to the R&S Spectrum Rider via the USB or the LAN interface. The number of
standards the R&S Spectrum Rider can store in its memory depends on the number of
other data sets stored on the R&S Spectrum Rider.
Performing Spectrum Measurements
www.allice.de