DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:PDIVision
.......................................... 404
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RPOSition
......................................... 404
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RLEVel
..............................................404
............................................................................................... 405
............................................................................................... 405
...............................................................................................406
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:PDIVision
<DevPerDiv>
This command sets the Y scale (deviation per division) in Hz (FM) or percent (AM).
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Parameters:
<DevPerDiv>
Numeric value with the unit range of Hz | kHz| MHz| MAHz| GHz
| PCT.
Default unit: GHz
Example:
DISP:TRAC:Y:PDIV 5 PCT
Set the Y scale (deviation per division) as 5 % in the AM modu-
lation.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RPOSition
<RefLvlPosition>
This command defines the position of the reference level on the display grid.
First, you have to set a logarithmic scale for the vertical axis with
WINDow]:TRACe<1...2>:Y:SPACing
on page 319. For a linear scale, you cannot
modify the reference position.
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Parameters:
<RefLvlPosition>
Defines the reference position in %.
Range:
0 to 10
*RST:
10
Example:
DISP:TRAC:Y:RPOS 5
Sets the reference position to the 5th grid line.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RLEVel
<RefLevel>
This command defines the reference level.