Remote Commands
R&S
®
Spectrum Rider FPH
443
User Manual 1321.1011.02 ─ 11
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
DISP:TRAC:MEM ON
Activates the memory trace.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:MODE
<DisplayMode>
This command selects the trace mode.
If you are using the average, max hold or min hold trace mode, you can set the number
of measurements with
[SENSe:]SWEep:COUNt<SweepCount>
. Note that synchroni-
zation to the end of the average count is possible only in single sweep mode.
The numeric suffix at DISPlay is irrelevant for this command.
The numeric suffix at TRACe selects the number of the trace.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Parameters:
<DisplayMode>
WRITe | AVERage | MINHold | MAXHold | VIEW | FREeze |
INFinite
*RST:
WRITe
You can turn off the trace with
DISPlay<1...2>[:WINDow]:TRACe<1...2>[:STATe]
.
For more information, see chapter "Trace Mode".
Example:
SWE:CONT OFF
SWE:COUN 16
Turn on single sweep mode and sets the number of measure-
ments to 16.
DISP:TRAC:MODE MAXH
Activates MAXHold mode for the trace.
INIT;*WAI
Starts the measurement and waits for the end of the 16 sweeps.
DISPlay[:WINDow]:TRACe:STYLe
<TraceStyle>
This command selects the style of the trace display.
Parameters:
<TraceStyle>
LINes | POLYgon
LINes
Adjacent Channel Leakage Ratio measurement (ACLR or ACP).
POLYgon
Shows the results in form of vertical lines at each receiver fre-
quency.
Remote Control - Commands