Remote Commands
R&S
®
Spectrum Rider FPH
387
User Manual 1321.1011.02 ─ 11
You can use an isotropic antenna with all measurement modes that are described
above.
List of commands
............................................................................................... 387
[SENSe:]CORRection:TRANsducer<1...2>:ISOTropic[:STATe]
............................................387
INPut:ANTenna:MEASure
<Direction>
This command selects the direction that the isotropic antenna measures.
Parameters:
<Direction>
AUTO | X | Y | Z
AUTO
Automatically selects the direction.
X | Y | Z
Measures the corresponding direction only.
*RST:
AUTO
Example:
INP:ANT:MEAS X
Measures in x-direction.
INPut:ANTenna:STATe
<State>
This command turns the use of an isotropic antenna on and off.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
INP:ANT:STAT ON
Activates the isotropic antenna.
[SENSe:]CORRection:TRANsducer<1...2>:ISOTropic[:STATe]
<State>
This command queries if the currently selected transducer is an isotropic antenna.
When you use the command for 3GPP WCDMA measurements in the Digital Modula-
tion operating mode, the command also turns the use of an isotropic antenna on and
off.
The numeric suffix at TRANsducer specifies the primary or secondary transducer.
Suffix:
<1...2>
.
1...2
Parameters:
<State>
ON | OFF
0
Transducer is not an isotropic antenna.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH