Getting Started
R&S
®
Spectrum Rider FPH
66
User Manual 1321.1011.02 ─ 11
2. Select either "Repetitive" or "Once" to turn on the audio signal.
If you have selected "Once", the R&S Spectrum Rider beeps once if the battery
runs out of power. For a continuous beep, select "Repetitive".
3. Select "Off" to turn off the beeper.
Auto power up
The R&S Spectrum Rider allows you to auto power up the instrument when supplying
with an AC power, without pressing the [POWER] key button.
1. In the "Instrument Setup" dialog box, select the "Auto Power Up on AC" menu item.
2. Select "On" to turn on the auto power up mode.
On the next power up cycle, the R&S Spectrum Rider goes into the auto power up
mode when supplying with an AC power.
3.2.7.9
Internal Alignment
R&S Spectrum Rider option
R&S FPH-K35 (order number: 1321.1563.02) option is required to operate the R&S
Spectrum Rider internal frequency clock and reference level adjustment.
Operating mode
The frequency clock and reference level alignment can only be performed in "Spec-
trum" mode. If wrong operating mode is detected, a warning message is displayed
before alignment is performed.
For more information on the operating mode of the R&S Spectrum Rider, see
Instrument Tour
www.allice.de