
Remote Commands
R&S
®
Spectrum Rider FPH
357
User Manual 1321.1011.02 ─ 11
This command is a query and therefore has no *RST value.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
LIMit selects the limit line.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Example:
CALC:LIM1:UNIT:X?
Queries the x-unit of the first limit line.
Usage:
Query only
CALCulate<1...2>:LIMit<1...2>:UNIT[:Y]?
This command queries the vertical unit of a limit line.
This command is a query and therefore has no *RST value.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
LIMit selects the limit line.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Example:
CALC:LIM1:UNIT?
Queries the y-unit of the first limit line.
Usage:
Query only
CALCulate<1...2>:LIMit<1...2>:LOWer:SELect
<LimitLine>
This command selects the lower limit line.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
LIMit selects the limit line.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Parameters:
<LimitLine>
String containing the filename of the lower limit line.
Example:
CALC:LIM1:LOW:SEL 'GSM_Lower.rellim'
Selects the lower limit line.
Usage:
Setting only
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH