Maintenance and Trouble Shooting
89
C
C
3523
Hepa filter fan
sensor failed.
Hepa filter fan sensor
malfunction.
Hepa filter fan sensor is not
working properly.
Restart system. If problem
persists, contact Technical
Service.
3524
Heater unit 1
sensor failed.
Heater unit 1 sensor
malfunction.
Heater unit 1 sensor is not
working properly.
Restart system. If problem
persists, contact Technical
Service.
3525
Heater unit 2
sensor failed.
Heater unit 2 sensor
malfunction.
Heater unit 2 sensor is not
working properly.
Restart system. If problem
persists, contact Technical
Service.
3526
Database
access error.
Connection to database
could not be established.
Connection to database
could not be established.
Restart system. If problem
persists, contact Technical
Service.
3534
Changes
cannot be
saved.
Changed list cannot be
saved.
Changes in the sample
material or operators list
cannot be saved.
Restart system. If problem
persists, contact Technical
Service.
3546
File Access
Error
Unable to save data of result
screen, teaching log or run
log.
Unable to save data of result
screen, teaching log or run
log.
3547
New database
was created.
New database was created.
Original database was not
found or not functioning
properly.
New database was created.
Original database was not
found or not functioning
properly.
3549
Insufficient
Sample
volume.
[Ch_X]
Smaller volume than
selected was detected in the
sample tube in referenced
channel number.
Check and add sample
volume in respective
channel, if needed.
Smaller volume than selec-
ted was detected in the
sample tube in referenced
channel number.
Check and add sample
volume in respective
channel, if needed, and
restart run.
3550
Sample
missing.
[Ch_X]
No sample was detected in
referenced channel number.
Check if sample tube has
been inserted and contains
the sample.
No sample was detected in
referenced channel number.
Check if sample tube has
been inserted and contains
the sample. Then restart
run.
3551
Sample
clot.[Ch_X]
A clot has been detected in
the sample of the referenced
channel number.
Check the sample for clots,
follow the pack insert
recommendation for
characteristics of sample.
A clot has been detected
in the sample of the
referenced channel number.
Check the sample for clots,
follow the pack insert
recommendation for
characteristics of sample.
Then restart run.
3552
Moving range
error
The range of possible
movements of the axis was
exceeded.
The range of possible
movements of the axis was
exceeded.
Restart system. If problem
persists, contact Technical
Service.
3553
There is no
correspon-
ding record
No data found in database
corresponding to entered
parameters.
No data found correspon-
ding to entered parameters.
Check search parameter
entries. If problem persists,
contact Technical Service.
Error
Code
ID
Error
Message
Error Information
(displayed on
information screen)
Explanation
Action
Error Codes and Trouble Shooting Guide
Instrument Error Codes
MagNA Pure Compact Manual.book Seite 89 Dienstag, 28. Februar 2006 4:16 16
Содержание MagNA Pure Compact Instrument
Страница 2: ......
Страница 6: ......
Страница 11: ......
Страница 35: ...www roche applied science com MagNA Pure Compact Operator s Manual Instrument Version 1 0 Software Version 1 1...
Страница 36: ......
Страница 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Страница 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Страница 95: ...Maintenance and Trouble Shooting MagNA Pure Compact Manual book Seite 69 Dienstag 28 Februar 2006 4 16 16...
Страница 129: ......
Страница 130: ...www roche applied science com 04555791001 0306 Roche Diagnostics GmbH Roche Applied Science 68298 Mannheim Germany...