MagNA Pure Compact Operator’s Manual - Version 1.3
16
A
A
2.4
Setup
2.4.1
Selecting a Location
The MagNA Pure Compact Instrument requires very little setup. Choose for the MagNA
Pure Compact Instrument a clean, dry, level, stable surface within 3 m of a compatible
electrical outlet. To ensure proper ventilation, leave 10 cm of space behind the instrument
and 15 cm at each side of the instrument. No space is needed at the back of the instru-
ment.
Protect the MagNA Pure Compact Instrument from heat and excessive sunlight and
always ventilate the room well. The instrument is for „indoor use only“ and should not be
operated in areas of excessive humidity or extremes of temperature. Do not use the
instrument where there is a risk of explosion.
2.4.2
Easy Teaching Procedure
The software has been preinstalled by the manufacturer. The installer (service user) must
perform an adjustment procedure, called Easy Teaching, before the instrument can be
used. Only a service operator, not a regular operator, can access the menu for the Easy
Teaching procedure. This is done by your local Roche Diagnostics representative.
Note:
The Easy Teaching Procedure must be performed before the instrument will work
correctly. If the adjustment is not done properly, either the instrument could be damaged
or the instrument performance could be affected (e.g. pipetting accuracy, yield in nucleic
acids, etc.).
Installation of the Instrument
Setup
MagNA Pure Compact Manual.book Seite 16 Dienstag, 28. Februar 2006 4:16 16
Содержание MagNA Pure Compact Instrument
Страница 2: ......
Страница 6: ......
Страница 11: ......
Страница 35: ...www roche applied science com MagNA Pure Compact Operator s Manual Instrument Version 1 0 Software Version 1 1...
Страница 36: ......
Страница 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Страница 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Страница 95: ...Maintenance and Trouble Shooting MagNA Pure Compact Manual book Seite 69 Dienstag 28 Februar 2006 4 16 16...
Страница 129: ......
Страница 130: ...www roche applied science com 04555791001 0306 Roche Diagnostics GmbH Roche Applied Science 68298 Mannheim Germany...