MagNA Pure Compact Operator’s Manual - Version 1.3
50
B
B
How to Operate the MagNA Pure Compact Instrument
B
B
For best results, grab the flap that is on the opposite end of the cartridge from the two isolated wells. Then, with the
two isolated wells pointing away from you, insert all the wells on the plastic cartridge into the holes in the cartridge
rack. Use the guide slots on the rack to help position the cartridge.
Note:
Make sure the cartridge fits all the way into the rack. Do not simply lay the cartridge on top of the
rack, since, if it sticks up, it may move during the run and cause the instrument to malfunction.
If you are going to process more than one sample
during the current purification run:
Repeat the above cartridge identification and place-
ment steps 4-5 for each prefilled cartridge until all
(up to 8) cartridges are recorded on screen and
inserted in the cartridge rack.
Result: On the overview diagram, the last added cart-
ridge position will be green (a), and the previously
added positions will be dark blue (b). The name of
the appropriate purification kit will also be displayed
(c).
Note:
It is not possible to enter barcodes of the same
type (catalogue number). Barcodes from identical or
expired cartridges are rejected.
When all cartridges have been programmed and
inserted into the cartridge rack, do the following:
䉴
Reinsert the Cartridge Rack into the MagNA Pure
Compact Instrument. To reinsert the Cartridge
Rack, fit the pin on the stage into the wide part of
the keyhole on the rack, then push the rack back,
which will also immobilize the pin in the narrow
part of the keyhole. Now you can lower the stage
by moving the handle away from you.
䉴
On SAMPLE ORDERING SCREEN 1, touch the
Cartridge Insertion confirmation button (a). After
you touch the confirmation button, the symbol on
the button changes from "?" to a checkmark.
䉴
The Next
button (b) will now be active.
Touch this button to confirm the information on
this screen and go to the next programming
screen.
Action
Handling
Software programming
5
6
7
Performing a Purification Run
Purification
Performing a Purification Run
Purification
51
4_MagNA Pure Compact_ChapterB_2.fm Seite 50 Mittwoch, 8. März 2006 10:13 10
Содержание MagNA Pure Compact Instrument
Страница 2: ......
Страница 6: ......
Страница 11: ......
Страница 35: ...www roche applied science com MagNA Pure Compact Operator s Manual Instrument Version 1 0 Software Version 1 1...
Страница 36: ......
Страница 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Страница 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Страница 95: ...Maintenance and Trouble Shooting MagNA Pure Compact Manual book Seite 69 Dienstag 28 Februar 2006 4 16 16...
Страница 129: ......
Страница 130: ...www roche applied science com 04555791001 0306 Roche Diagnostics GmbH Roche Applied Science 68298 Mannheim Germany...