MagNA Pure Compact Operator’s Manual - Version 1.3
52
B
B
How to Operate the MagNA Pure Compact Instrument
B
B
SAMPLE ORDERING SCREEN 2 appears. While at this screen, do the following:
䉴
In the Protocol field, either accept the protocol (from the last run) that appears in the field or select another puri-
fication protocol from the adjacent pull-down menu. (Touch the Down arrow to the right of the field to access
this menu.)
䉴
In the Sample Volume field, select a value from the adjacent pull-down menu.
䉴
In the Sampe Material field, select your sample
material from the adjacent pull-down menu.
䉴
In the Elution Volume field, select a volume from
the adjacent pull-down menu.
Note:
For an overview about the possible
combinations of different sample and
elution volumes, please refer to the
respective pack insert of the purification kit.
Combining of non recommended volumes
may leed to suboptimal results (
e.g.
processing performance, yield, etc.)
䉴
(optional) If you are using an internal control,
select a control volume from the pull-down menu
adjacent to the Internal Control Volume field:
You can choose between 5, 10, or 20 µl. Default
setting is None. Note, that some purification
protocols do not allow an optional internal
control.
.
䉴
Insert the appropriate number of Tip Trays (one
per purification) in the instrument Tip Rack.
䉴
On SAMPLE ORDERING SCREEN 2, touch the
"Tip Trays inserted" confirmation button (1).
Result:
䉴
After you touch the "Tip Trays inserted" confirma-
tion button, the symbol on the button changes
from "?“ to a checkmark (2). Also, the reagent
layout diagram shows the tip trays inserted (in
dark blue).
䉴
The Next button
will now be active. Touch
this button to confirm the information on this
screen and go to the next programming screen.
Action
Handling
Software programming
8
Performing a Purification Run
Purification
Performing a Purification Run
Purification
53
4_MagNA Pure Compact_ChapterB_2.fm Seite 52 Mittwoch, 8. März 2006 10:13 10
Содержание MagNA Pure Compact Instrument
Страница 2: ......
Страница 6: ......
Страница 11: ......
Страница 35: ...www roche applied science com MagNA Pure Compact Operator s Manual Instrument Version 1 0 Software Version 1 1...
Страница 36: ......
Страница 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Страница 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Страница 95: ...Maintenance and Trouble Shooting MagNA Pure Compact Manual book Seite 69 Dienstag 28 Februar 2006 4 16 16...
Страница 129: ......
Страница 130: ...www roche applied science com 04555791001 0306 Roche Diagnostics GmbH Roche Applied Science 68298 Mannheim Germany...