MagNA Pure Compact Operator’s Manual - Version 1.3
82
C
C
2.1
Counter and Reminder
The MagNA Pure Compact Instrument Software has a Counter and Reminder function
which counts the number of runs, number of hours of UV lamp usage, and number of
hours of operation.
Your Roche Diagnostics service representative is able to set reminders based on these
individual counters or on a fixed date e.g. to be reminded for the regular O-ring or UV
lamp exchange. If the counter reaches the set "Next Reminder" value the reminder text
defined is displayed on the screen.
To view the current counter and set reminder values, you can access the Counter and
Reminder Screen via the respective button on the Maintenance Menu Screen. Reminder
values can changed by the service representative only.
2.2
Remote Control and Host
Your Roche service representative is able to perform some of the service and trouble-
shooting via the remote control function. For this purpose there are two buttons Remote
Control and Host available in the Maintenance Menu. In order to get support via remote
control, please contact directly your local service representative for enabling this option.
Service Maintenance
Counter and Reminder
MagNA Pure Compact Manual.book Seite 82 Dienstag, 28. Februar 2006 4:16 16
Содержание MagNA Pure Compact Instrument
Страница 2: ......
Страница 6: ......
Страница 11: ......
Страница 35: ...www roche applied science com MagNA Pure Compact Operator s Manual Instrument Version 1 0 Software Version 1 1...
Страница 36: ......
Страница 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Страница 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Страница 95: ...Maintenance and Trouble Shooting MagNA Pure Compact Manual book Seite 69 Dienstag 28 Februar 2006 4 16 16...
Страница 129: ......
Страница 130: ...www roche applied science com 04555791001 0306 Roche Diagnostics GmbH Roche Applied Science 68298 Mannheim Germany...