MagNA Pure Compact Operator’s Manual - Version 1.3
54
B
B
How to Operate the MagNA Pure Compact Instrument
B
B
Do one of the following:
SAMPLE ORDERING SCREEN 3 appears. While at
this screen, do the following:
䉴
Pipet the correct type and amount of sample
material (as chosen on Sample Ordering Screen
2) into a sample tube (tube without barcode label,
from the reagent kit).
䉴
Identify the tube by attaching your own barcode
label or writing a number on the tube with a per-
manent marker.
䉴
Put the filled sample tube in position 1 on row 1
of the Tube Rack.
Note:
It is recommended to use the
appropriate row in the tube rack for original
sample tubes, if suitable for these tubes
(e.g. max. 15 ml Falcon tubes).
䉴
On SAMPLE ORDERING SCREEN 3, touch Sample ID field 1. In that field, enter the ID of Sample 1 (i.e., the one
you just put in the rack) by either scanning the (self-attached) barcode or using the virtual keyboard to type the
ID number. (Touch the Keyboard button to access the keyboard.)
䉴
(optional) In the Comment field, you may enter
additional information about Sample 1. (Touch the
field to highlight it, then use the virtual keyboard.)
䉴
Repeat the above sample placement and
identification steps for each sample to be purified
(e.g., for Sample position 2, 3, …, 8).
Result: As you fill in the "Sample-ID" field for each
sample, the corresponding position in the reagent
layout will change from green (active) to dark blue
(installed in instrument).
Action
Handling
Software programming
9
If ...
Then ...
You are not using internal
controls (i.e., you selected
a protocol with no internal
control on SAMPLE
ORDERING SCREEN 2)
(Action 8 above)
䉴
Insert the tube rack into the MagNA Pure
Compact Instrument.
䉴
Touch the "Samples inserted" confirmation
button (a).
䉴
The Next button (b) will now be active. Touch
this button to confirm the information on this
screen and go to the next programming
screen (SAMPLE ORDERING SCREEN 5).
䉴
Go to Action 11 below.
You are using internal
controls (i.e., you selected
a protocol with internal
control on SAMPLE
ORDERING SCREEN 2)
䉴
Touch the "Sample inserted" confirmation
button (a).
䉴
The Next button (b) will now be active. Touch
this button to confirm the information on this
screen and go to the next programming
screen (SAMPLE ORDERING SCREEN 4).
䉴
Go to Action 10 below.
Performing a Purification Run
Purification
Performing a Purification Run
Purification
55
4_MagNA Pure Compact_ChapterB_2.fm Seite 54 Mittwoch, 8. März 2006 10:13 10
Содержание MagNA Pure Compact Instrument
Страница 2: ......
Страница 6: ......
Страница 11: ......
Страница 35: ...www roche applied science com MagNA Pure Compact Operator s Manual Instrument Version 1 0 Software Version 1 1...
Страница 36: ......
Страница 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Страница 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Страница 95: ...Maintenance and Trouble Shooting MagNA Pure Compact Manual book Seite 69 Dienstag 28 Februar 2006 4 16 16...
Страница 129: ......
Страница 130: ...www roche applied science com 04555791001 0306 Roche Diagnostics GmbH Roche Applied Science 68298 Mannheim Germany...