With the O-Ring Exchange Tool, remove the O-rings as shown below:
Place the pin of the O-Ring Exchange Tool
into the hole of a nozzle:
Insert the pin into the nozzle as deeply as
possible.
While maintaining the position of the pin,
move the O-ring pusher toward the pin
until it is inserted into the groove in the
nozzle.
Move the O-ring pusher closer to the pin
until it displaces the O-ring from the
groove.
Use your thumb to pull down the part of
the O-ring that now sticks out of the
groove. Roll the O-ring down around the
shaft of the Exchange Tool.
Relax your forefinger until the O-ring
pusher pops out of the nozzle groove.
1
2
3
4
5
6
User Maintenance
O-ring Maintenance
MagNA Pure Compact Manual.book Seite 78 Dienstag, 28. Februar 2006 4:16 16