![Roche MagNA Pure Compact Instrument Скачать руководство пользователя страница 121](http://html1.mh-extra.com/html/roche/magna-pure-compact-instrument/magna-pure-compact-instrument_operators-manual_1472301121.webp)
Maintenance and Trouble Shooting
95
C
C
Poor (RT-) PCR
performance /
poor sensitivity
Too much nucleic acid in PCR
Check nucleic acid concentration in the eluted sam-
ples and adjust amount of eluate per PCR. Optimum:
e.g. for DNA 10-100 ng/PCR.
Use less starting material. Refer to the respective pack
insert of the reagent kit.
Not enough nucleic acid in PCR
Check nucleic acid concentration in the eluted sam-
ples and adjust amount of eluate per PCR. Optimum:
e.g. for DNA 10–100 ng/PCR.
Use more starting material, refer to the respective pack
insert of the reagent kit.
Poor purity of nucleic acid
Too much sample material, refer to the respective pack
insert of the reagent kit.
Storage of samples was not optimal Use fresh or frozen samples; avoid using samples that
were stored at room temperature for extended periods
of time.
Improper storage of eluate
Do not store eluates at room temperature;
do not perform overnight runs.
Incorrect anticoagulant was used
Use EDTA or citrated blood or plasma. Do not use
heparinized blood or plasma.
(RT-) PCR reagents and protocols
are not optimal
Check (RT-) PCR reagents and protocols with a posi-
tive DNA or RNA control (
e.g.
, Human Genomic DNA,
Cat. No. 1 691 112) or use an internal control.
Problem
Possible cause
Recommendation
Error Codes and Trouble Shooting Guide
Trouble Shooting Guide
MagNA Pure Compact Manual.book Seite 95 Dienstag, 28. Februar 2006 4:16 16
Содержание MagNA Pure Compact Instrument
Страница 2: ......
Страница 6: ......
Страница 11: ......
Страница 35: ...www roche applied science com MagNA Pure Compact Operator s Manual Instrument Version 1 0 Software Version 1 1...
Страница 36: ......
Страница 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Страница 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Страница 95: ...Maintenance and Trouble Shooting MagNA Pure Compact Manual book Seite 69 Dienstag 28 Februar 2006 4 16 16...
Страница 129: ......
Страница 130: ...www roche applied science com 04555791001 0306 Roche Diagnostics GmbH Roche Applied Science 68298 Mannheim Germany...