Maintenance and Trouble Shooting
93
C
C
Errors during Self Test (in pop-up windows)
5601
Door is open
(5601).
Retry?
_
Instrument Door was open
during initial self test. Door
must be shut.
Close Door tightly and
press OK in message box.
5602
Door cannot
be lokked
(5602).
Retry?
_
Instrument door cannot be
locked during self test of
instrument. Door must be
locked during nozzle head
movement.
Close Door tightly and
press OK in message box.
5603
Crash sensor
failed (5603).
Retry?
_
Internal error regarding
crash sensor. Sensor failed
during self test.
Press OK for retry. If that
fails restart system. If
problem persists, contact
Technical Service.
5604
Pressure
sensor failed
(5604).
Retry?
_
Internal error regarding
pressure sensor. Sensor
failed during self test.
Press OK for retry. If that
fails restart system. If
problem persists, contact
Technical Service.
5605
Magnetic
home sensor
failed (5605).
Retry?
_
Internal error regarding
magnetic home sensor.
Sensor failed during self
test.
Press OK for retry. If that
fails restart system. If
problem persists, contact
Technical Service.
5606
Magnetic
sensor failed
(5606).
Retry?
_
Internal error regarding
magnetic sensor. Sensor
failed during self test.
Press OK for retry. If that
fails restart system. If
problem persists, contact
Technical Service.
5607
Z axis sensor
failed (5607).
Retry?
_
Internal error regarding Z
axis sensor. Sensor failed
during self test.
Press OK for retry. If that
fails restart system. If
problem persists, contact
Technical Service.
5608
M axis sensor
failed (5608).
Retry?
_
Internal error regarding M
axis sensor. Sensor failed
during self test.
Press OK for retry. If that
fails restart system. If
problem persists, contact
Technical Service.
5609
Y axis sensor
failed (5609).
Retry?
_
Internal error regarding Y
axis sensor. Sensor failed
during self test.
Press OK for retry. If that
fails restart system. If
problem persists, contact
Technical Service.
Error
Code
ID
Error
Message
Error Information
(displayed on
information screen)
Explanation
Action
Error Codes and Trouble Shooting Guide
Instrument Error Codes
MagNA Pure Compact Manual.book Seite 93 Dienstag, 28. Februar 2006 4:16 16
Содержание MagNA Pure Compact Instrument
Страница 2: ......
Страница 6: ......
Страница 11: ......
Страница 35: ...www roche applied science com MagNA Pure Compact Operator s Manual Instrument Version 1 0 Software Version 1 1...
Страница 36: ......
Страница 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Страница 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Страница 95: ...Maintenance and Trouble Shooting MagNA Pure Compact Manual book Seite 69 Dienstag 28 Februar 2006 4 16 16...
Страница 129: ......
Страница 130: ...www roche applied science com 04555791001 0306 Roche Diagnostics GmbH Roche Applied Science 68298 Mannheim Germany...