MagNA Pure Compact Operator’s Manual - Version 1.3
90
C
C
3555
No protocol
files found
No valid protocol for the
entered cartridge ID found.
No valid protocol for the
entered cartridge ID found.
Restart sample ordering
and enter cartridge ID
again. Install appropriate
protocol, if available. If
problem persists, contact
Technical Service.
3556
Exe file could
not be found
The file that was specified by
the ini file does not exist.
When remote access button
was pressed:
The executable file that was
configured for remote
access could not be found.
If remote access is desired:
Restart system. If problem
persists, contact Technical
Service.
3557
Empty waste
tank
Waste tank has not been
emptied after Liquid Waste
Discard.
After Liquid Waste discard
process the waste tank has
to be emptied. This is
controlled by a waste tank
sensor.
Empty waste tank.
3558
Folder does
not exist
The specified folder be
found. Files can not be
saved.
The save folder that is set in
[Setup>Settings] could not
be found.
Confirm existence of folder
in the [Setup>Settings]
screen.
3559
Could not
create folder
The save folder was not
existent and could not be
created. Files can not be
saved.
The save folder that is set in
[Setup>Settings] could not
be found and could not be
created.
Confirm existence of folder
or drive in the [Setup>Set-
tings] screen.
3560
New folder
was created
The save folder was not
existent and was created.
The save folder did not exist
and was created.
3561
No Cartridge
[Ch_X]
Ordered cartridge was not
found.
A cartridge that was orde-
red during sample ordering
was not found.
Check stage setup.
3562
Temperature
Error [Ch_X]
Temperature that was called
for by the chosen protocol
could not be reached.
Temperature that was cal-
led for by the chosen proto-
col could not be reached.
Restart system. If problem
persists, contact Technical
Service.
3563
Unexpected
Error [Ch_X]
An unexpected error has
occured.
An unexpected error has
occured.
Restart system. If problem
persists, contact Technical
Service.
3564
Waste tank is
not inserted
The waste is not inserted
into the instrument.
The waste is not inserted
into the instrument. The
instrument will not perform
Liquid Waste Discard wit-
hout a waste tank.
Insert waste tank.
3566
Hepa Filter
Fan stopped
Hepa fan does not work. The
Run function is suspended.
Hepa filter fan does not
work. The Run function is
suspended
Restart system. If problem
persists, contact Technical
Service.
3568
Plausibility test
failed
The plausibility test of axes
and sensors failed.
Check stage and restart the
system. If problem persists
contact Technical Service.
The instrument tests axes
and sensors with a
plausibility test which failed.
Check stage and restart
the system. If problem
persists contact Technical
Service.
Error
Code
ID
Error
Message
Error Information
(displayed on
information screen)
Explanation
Action
Error Codes and Trouble Shooting Guide
Instrument Error Codes
MagNA Pure Compact Manual.book Seite 90 Dienstag, 28. Februar 2006 4:16 16
Содержание MagNA Pure Compact Instrument
Страница 2: ......
Страница 6: ......
Страница 11: ......
Страница 35: ...www roche applied science com MagNA Pure Compact Operator s Manual Instrument Version 1 0 Software Version 1 1...
Страница 36: ......
Страница 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Страница 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Страница 95: ...Maintenance and Trouble Shooting MagNA Pure Compact Manual book Seite 69 Dienstag 28 Februar 2006 4 16 16...
Страница 129: ......
Страница 130: ...www roche applied science com 04555791001 0306 Roche Diagnostics GmbH Roche Applied Science 68298 Mannheim Germany...