Maintenance and Trouble Shooting
79
C
C
Grease the new O-rings as described in Section C.1.2.1 above.
Note:
Lubricating the O-rings before placing them on the nozzles is important to ensure
correct reaction tip uptake and alignment.
䉴
Install new O-rings as shown below:
While holding the O-ring on the Exchange
Tool with your thumb, pull the entire O-
ring changer downward, away from the
nozzle.
The O-ring is now removed from the
nozzle:
Place a new, greased O-ring on the shaft
of the O-Ring Exchange Tool. Make sure
the O-ring is well below the pin.
Insert the pin of the O-Ring Exchange
Tool as deep as possible into the nozzle
hole:
Use your thumb and forefinger to push
the O-ring up the shaft onto the nozzle.
Continue pushing until the O-ring is
seated in the nozzle groove.
Slowly withdraw the Exchange Tool from
the nozzle.
7
8
1
2
3
User Maintenance
O-ring Maintenance
MagNA Pure Compact Manual.book Seite 79 Dienstag, 28. Februar 2006 4:16 16
Содержание MagNA Pure Compact Instrument
Страница 2: ......
Страница 6: ......
Страница 11: ......
Страница 35: ...www roche applied science com MagNA Pure Compact Operator s Manual Instrument Version 1 0 Software Version 1 1...
Страница 36: ......
Страница 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Страница 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Страница 95: ...Maintenance and Trouble Shooting MagNA Pure Compact Manual book Seite 69 Dienstag 28 Februar 2006 4 16 16...
Страница 129: ......
Страница 130: ...www roche applied science com 04555791001 0306 Roche Diagnostics GmbH Roche Applied Science 68298 Mannheim Germany...