67 • Maintenance
03 Series Operator’s Manual
5. Remove the front case from the rear case.
Figure 15: Case Separation
6. Carefully remove the old sensor from the sensor socket.
7. Carefully insert the replacement sensor in the socket. Make sure the
sensor face with the colored ring is facing up and that the sensor is
keyed correctly (HS-03 and CO-03 only).
Figure 16: Replacing the Sensor
CAUTION: When replacing the sensor, verify that the sensor is properly
aligned with its socket before inserting it into the socket. The
CO and H
2
S sensors have alignment slots which match up
with alignment tabs in the sockets. Forcing a sensor into its
socket may damage the sensor or the socket.
Front Case
Rear Case
Screw, 4x
Tabs
Oxygen
Sensor
H2S
Sensor
CO
Sensor
Align Slots in H2S
and CO Sensor
with Tabs in Case
Oxygen
Sensor Not
Keyed
Rear
Case