33 • Calibration Mode
03 Series Operator’s Manual
Setting the Date and Time
Entering the DATE menu item allows you to set the date and time.
1. When the DATE Screen is displayed, press and release the POWER
MODE button. A screen appears with the year flashing in the upper left,
the month and day in the middle, and the time in the lower right.
2. Use the AIR button to display the desired year.
3. Press and release the POWER MODE button to save the setting. The
month setting flashes.
4. Repeat steps 2 and 3 to enter the month, day, hour and minute setting.
When you save the minute setting, END is displayed before the
instrument returns to the DATE Screen.
Performing a Fresh Air Adjustment
Entering the AIR menu item allows you to perform a fresh air adjustment.
Perform a fresh air adjustment in Calibration Mode when you are
performing a calibration before proceeding either to the A--CAL or M--CAL
menu item to perform a span adjustment (zero adjustment for OX-03). A
fresh air adjustment performed in Calibration Mode is the same as a fresh
air adjustment in Normal Mode. The AIR menu item is available in
Calibration Mode for convenience when performing a complete calibration.
WARNING: Calibrate the 03 Series in a non-hazardous environment.
DATE
2014
2.11
3:05
AIR