03 Series Operator’s Manual
Calibration Mode • 50
4. Confirm that the regulator on/off knob is turned all the way clockwise
(closed) and screw the calibration gas cylinder onto the regulator.
5. Push the calibration adapter onto the 03 Series’ sensor face as shown
in Figure 12. The calibration adapter secures itself to the instrument by
latching on to two recesses in the instrument’s rear case. The rubber
boot (if installed) does not need to be removed to install the adapter.
Figure 12: Installing the Calibration Adapter
6. Press and release the POWER MODE button. The LCD will display the
current gas reading, “BUMP” and “APPLY” will alternate at the top of the
screen, and the instrument will begin a count down whose length
depends on how the Bump Test Time(sec) parameter in the 03 Series
User Setup Program is set. The factory setting is 30 seconds.
7. Turn the regulator on/off knob counterclockwise to open it. Calibration
gas will begin to flow. Do not delay opening the regulator.
8. When the bump test countdown reaches 0, the unit will determine the
bump test results and proceed as described in either Step 9 or Step 10
depending on the setting of Calibration After Bump Test Failed. See
POW ER
MO DE
CO-03
Calibration
Adapter
AIR
APPLY
ppm
15
30
BUMP
ppm
15
30