03 Series Operator’s Manual
Start Up • 10
Start Up
This section explains how to start up the 03 Series and to get it ready for
operation.
NOTE:
The screens shown in this section are for a CO-03. If you are
using a different version of the 03 Series, your instrument screen
will vary slightly from those shown below.
Start-up Procedure
1. Press and briefly hold down the POWER MODE button. The backlight
will turn on and all the display segments will turn on. Release the button
when you hear a beep.
2. The vibrator vibrates and the alarm light flashes momentarily.
3. The initial startup screen depends on how Bump Fail Behavior is set.
This parameter along with the Cal. Limit Display and Cal. Limit Check
parameters mentioned in Step 5 and the Auto Zero Adjustment
parameter mentioned in Step 12 below cannot be set using the 03
Series’ instrument menus, but are set using the 03 Series User Setup
Program. See the 03 Series User Setup Program Operator’s Manual for
information regarding changing various instrument parameters that are
not available for adjustment in the instrument’s operating modes.
•
If Bump Fail Behavior is set to None (factory setting) or if it is set to
Can Not Use
and the most recently performed bump test passed,
proceed to Step 5.
•
If Bump Fail Behavior is set to Can Not Use and the most recently
performed bump test failed, the following screen will appear.
The instrument cannot be used until a successful bump test or
calibration is performed. See “Calibration Mode” on page 31 for
instructions.
4. If Cal. Limit Display is set to Off, proceed to Step 6.
C.AL
FAIL