03 Series Operator’s Manual
Maintenance • 66
4. Carefully remove the old alkaline batteries.
NOTE:
To avoid having to reset the date and time, new batteries must be
installed within 2 minutes.
5. Carefully install the new AAA alkaline batteries. Follow the battery
diagram inside the battery compartment.
6. Reinstall the battery cover.
7. If the instrument has been without batteries for more than 2 minutes,
the date and time are reset and need to be set again. When the new
batteries are installed, the instrument will turn on automatically and will
display the Date/Time Screen. Set the date and time as described in
“Setting the Date and Time” on page 33. Once the date and time have
been set, the instrument will begin its warmup sequence. If you do not
set the date and time within 30 seconds, the instrument will
automatically begin its warmup sequence.
Replacing the Sensor
WARNING: Replace the sensor in a non-hazardous environment.
NOTE:
An alligator or belt clip may be installed on the instrument but is
not shown in the figures in this section.
1. Verify that the 03 Series is off.
2. Remove the rubber boot, if installed.
3. With a small Phillips screwdriver, carefully unscrew the four screws that
attach the rear case to the front case.
4. Turn the instrument right side up. The screws are not captive and will
fall out. Be sure not to lose them.