03 Series Operator’s Manual
Calibration Mode • 42
3. Press and release the POWER MODE button. The LCD will display the
current gas readings and “E--CAL.” will flash.
If you want to cancel the calibration, press and hold the AIR key for 2
seconds. The instrument will return to the E--CAL. Screen. If you hold
the AIR key down longer than 2 seconds, the instrument will start
scrolling through the Calibration Mode items.
4. Use the sample tubing to connect the calibration adapter to the
regulator. Attach the tubing to the calibration adapter on the inlet side as
shown below in Figure 7.
Figure 7: Calibration Kit Assembly
5. Confirm that the regulator on/off knob is turned all the way clockwise
(closed) and screw the calibration gas cylinder onto the regulator.
G--IN
E--CAL.
ppm
0
To Fixed Flow
Regulator
Calibration
Adapter
Calibration Tubing