03 Series Operator’s Manual
Operation • 22
5. Press and release the POWER MODE button again to proceed to the
TWA Screen. The TWA reading will be displayed.
6. Press and release the POWER MODE button again to proceed to the
Full Scale Screen. The detection range full scale value is displayed. An
“L” or an “A” will appear under the full scale value to indicate whether
the alarms are set to latching (L) or auto-resetting (A). This setting can
be adjusted using the 03 Series User Setup Program.
7. Press and release the POWER MODE button again to proceed to the
Date/Time Screen. The date and time of the instrument is displayed.
8. Press and release the POWER MODE button once again to return the
unit to Measuring Mode.
NOTE:
If you do not press a button for 20 seconds while in Display Mode,
the unit will return to Measuring Mode automatically and the
backlight will turn off.
Display Mode, OX-03
This section describes Display Mode for the OX-03 only. See “Display
Mode, CO-03 and HS-03” on page 20 for a description of the CO-03 and
HS-03 instruments’ Display Mode.
You can access Display Mode while in Measuring Mode by using the
POWER MODE button. In Display Mode you can view and reset the
minimum and maximum readings, view the detection range full scale, and
view the date and time.