03 Series Operator’s Manual
Calibration Mode • 52
•
To view the bump test gas reading, press the AIR button.
•
To return to the BUMP screen in Calibration Mode, press POWER
MODE at any time.
•
Turn the regulator on/off knob clockwise to close it.
If the instrument fails the bump test,
•
A calibration is immediately and automatically started. The
instrument will begin a countdown whose length depends on the
Bump Test Time(sec)
and Calibration Time(sec) After Bump
Test Failed
parameters’ values. These parameters can be changed
using the 03 Series User Setup Program. The countdown time for
the calibration is the Calibration Time(sec) After Bump Test
Failed
value (factory setting is 90 seconds) minus the Bump Test
Time(sec)
value (factory setting is 30 seconds). Continue to apply
the calibration gas.
BUMP
ppm
45
BUMP
APPLY
ppm
45
60
CAL.
ppm
45
60