03 Series Operator’s Manual
Calibration Mode • 34
1. Find a fresh air environment, an environment of normal oxygen content
(20.9%) that is free of toxic and combustible gasses.
2. When the AIR screen is displayed, press and release the POWER
MODE button. A screen appears that displays the current gas reading.
If you want to cancel the fresh air adjustment, press and release the
AIR button to return to the AIR screen. To continue with the fresh air
adjustment, proceed to Step 3.
3. Press and hold the AIR button. The LCD prompts you to continue to
hold the AIR button.
4. Release the AIR button when the following screen appears.
5. The unit will take a few seconds to adjust the fresh air reading, then
indicate “END” in the upper left corner and return to the AIR Screen.
Performing an Automatic Span Adjustment
(Zero Adjustment for OX-03) in A--CAL
The A--CAL menu item only appears in Calibration Mode if the One Touch
Cal Time
parameter in the 03 Series User Setup Program is set to 0
(factory setting). If One Touch Cal Time is set to anything other than 0, E--
CAL will replace A--CAL.
AIR
ppm
0
CO
AIR
HOLD
- - - -
ADJ
- - - -